A scattering filter for energy-dispersive optics

被引:6
作者
Hagelstein, M
Lienert, U
Ressler, T
San Miguel, A
Freund, A
Cunis, S
Schulze, C
Fontaine, A
Hodeau, JL
机构
[1] European Synchrotron Radiat Facil, F-38043 Grenoble, France
[2] Univ Ulm, D-89081 Ulm, Germany
[3] Paul Scherrer Inst, CH-5232 Villigen, Switzerland
[4] CNRS, F-38042 Grenoble, France
来源
JOURNAL OF SYNCHROTRON RADIATION | 1998年 / 5卷
关键词
energy-dispersive XAFS; Bonse-Hart camera; SAXS filters;
D O I
10.1107/S0909049597015252
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A.filtering technique to remove parasitic scattering from X-ray absorption spectra that are acquired in energy-dispersive mode has been developed and tested at the European Synchrotron Radiation Facility. The improved set-up removes small-angle scattering of the sample or the windows of sample cells which may spoil the energy resolution or reduce the intensity of prominent features in the absorption spectrum, such as the white line at the Pt L-III edge. The sample is placed behind the curved monochromator and between two plane perfect crystals in the Bonse-Hart configuration. The dispersion of the Bonse-Hart double-crystal camera is matched to the dispersion of the curved monochromator by inclining the scattering planes of the two optical elements against each other.
引用
收藏
页码:753 / 755
页数:3
相关论文
共 6 条
[1]   TAILLESS X-RAY SINGLE-CRYSTAL REFLECTION CURVES OBTAINED BY MULTIPLE REFLECTION - (SI GE CRYSTALS - E/T) [J].
BONSE, U ;
HART, M .
APPLIED PHYSICS LETTERS, 1965, 7 (09) :238-&
[2]   SMALL ANGLE X-RAY SCATTERING BY SPHERICAL PARTICLES OF POLYSTYRENE AND POLYVINYLTOLUENE [J].
BONSE, U ;
HART, M .
ZEITSCHRIFT FUR PHYSIK, 1966, 189 (02) :151-&
[3]   ON THE USE OF FOCUSING FOR SMALL-ANGLE SCATTERING EXPERIMENTS [J].
FREUND, AK .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 216 (1-2) :269-274
[4]  
Guinier A, 1963, X-ray Diffraction
[5]   The beamline ID24 at ESRF for energy dispersive X-ray absorption spectroscopy [J].
Hagelstein, M ;
San Miguel, A ;
Fontaine, A ;
Goulon, J .
JOURNAL DE PHYSIQUE IV, 1997, 7 (C2) :303-308
[6]   CURVED CRYSTAL TRANSMISSION OPTICS FOR ENERGY-DISPERSIVE X-RAY-ABSORPTION SPECTROSCOPY [J].
HAGELSTEIN, M ;
FERRERO, C ;
HATJE, U ;
RESSLER, T ;
METZ, W .
JOURNAL OF SYNCHROTRON RADIATION, 1995, 2 :174-180