Estimation of the strength of adhesion between a thermoresponsive polymer coating and nitinol wire

被引:29
作者
Burke, Martin [1 ]
Clarke, Brenda [1 ]
Rochev, Yuri [1 ]
Gorelov, Alexandar [2 ]
Carroll, William [1 ]
机构
[1] Natl Univ Ireland, Natl Ctr Biomed Engn Sci, Galway, Ireland
[2] Univ Coll Dublin, Dept Chem, Dublin 2, Ireland
关键词
D O I
10.1007/s10856-007-3274-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
As polymer coatings become more widely used in the biomedical device industry, both to improve biocompatibility and as coatings for localised drug delivery, quantitative methods to measure the adhesive strength between coatings and substrates become a very important consideration. The aim of this study was to take a method for estimating the interfacial fracture toughness of a film to a flat substrate and apply it to Nitinol wires used in the production of medical devices. An investigation into the affect of surface roughness on the fracture toughness was also conducted. For the present study, a thermoresponsive based Poly (N-isopropylacrylamide) polymer was coated onto nitinol wire substrates and the adhesion strength between the polymer and wire was measured using a nanoindentation technique. Different surface treated nitinol wires, with different surface topography and roughness were used, and the affect of these surface properties on adhesion strength was investigated. Results showed that it was possible to apply the delamination technique to wire samples and obtain fracture toughness values. Results also showed that the surface roughness is an important parameter that can affect the adhesion between a coating and the substrate. It was found that, as the average surface roughness increased so also did the adhesive strength between the coating and wire sample.
引用
收藏
页码:1971 / 1979
页数:9
相关论文
共 31 条
[1]
Poly(N-isopropylacrylamide) co-polymer films as potential vehicles for delivery of an antimitotic agent to vascular smooth muscle cells [J].
Doorty, KB ;
Golubeva, TA ;
Gorelov, AV ;
Rochev, YA ;
Allen, LT ;
Dawson, KA ;
Gallagher, WM ;
Keenan, AK .
CARDIOVASCULAR PATHOLOGY, 2003, 12 (02) :105-110
[2]
Evaluation of a new controlled-drug delivery concept based on the use of thermoresponsive polymers [J].
Eeckman, F ;
Moës, AJ ;
Amighi, K .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2002, 241 (01) :113-125
[3]
Fracture study of organic-inorganic coatings using nanoindentation technique [J].
Etienne-Calas, S ;
Duri, A ;
Etienne, P .
JOURNAL OF NON-CRYSTALLINE SOLIDS, 2004, 344 (1-2) :60-65
[4]
The effect of coating cracking on the indentation response of thin hard-coated systems [J].
Hainsworth, SV ;
McGurk, MR ;
Page, TF .
SURFACE & COATINGS TECHNOLOGY, 1998, 102 (1-2) :97-107
[5]
Local drug delivery in restenosis injury: thermoresponsive co-polymers as potential drug delivery systems [J].
Kavanagh, CA ;
Rochev, YA ;
Gallagher, WA ;
Dawson, KA ;
Keenan, AK .
PHARMACOLOGY & THERAPEUTICS, 2004, 102 (01) :1-15
[6]
Nanomechanical fracture-testing of thin films [J].
Kriese, MD ;
Boismier, DA ;
Moody, NR ;
Gerberich, WW .
ENGINEERING FRACTURE MECHANICS, 1998, 61 (01) :1-20
[7]
Lewis AL, 2004, BIO-MED MATER ENG, V14, P355
[8]
Interfacial fracture property determination of coated systems: A finite element study [J].
Li, HQ ;
Cai, X ;
Cheng, QL .
JOURNAL OF MATERIALS SCIENCE LETTERS, 2001, 20 (23) :2167-2171
[9]
Indentation deformation and fracture of thin polystyrene films [J].
Li, M ;
Palacio, ML ;
Carter, CB ;
Gerberich, WW .
THIN SOLID FILMS, 2002, 416 (1-2) :174-183
[10]
Fracture mechanisms of thin amorphous carbon films in nanoindentation [J].
Li, XD ;
Diao, DF ;
Bhushan, B .
ACTA MATERIALIA, 1997, 45 (11) :4453-4461