Direct detection of glucose by surface plasmon resonance with bacterial glucose/galactose-binding protein

被引:127
作者
Hsieh, HV [1 ]
Pfeiffer, ZA [1 ]
Amiss, TJ [1 ]
Sherman, DB [1 ]
Pitner, JB [1 ]
机构
[1] BD Technol, Biosense, Res Triangle Pk, NC 27709 USA
关键词
biosensor; continuous glucose monitoring; reagentless sensor; protein engineering; SPR; surface plasmon resonance;
D O I
10.1016/S0956-5663(03)00271-9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The monitoring and management of blood glucose levels are key components for maintaining the health of people with diabetes. Traditionally, glucose monitoring has been based on indirect detection using electrochemistry and enzymes such as glucose oxidase or glucose dehydrogenase. Here, we demonstrate direct detection of glucose using a surface plasmon resonance (SPR) biosensor. By site-specifically and covalently attaching a known receptor for glucose, the glucose/galactose-binding protein (GGBP), to the SPR surface, we were able to detect glucose binding and determine equilibrium binding constants. The site-specific coupling was accomplished by mutation of single amino acids on GGBP to cysteine and subsequent thiol conjugation. The resulting SPR surfaces had glucose-specific binding properties consistent with known properties of GGBP. Further modifications were introduced to weaken GGBP-binding affinity to more closely match physiologically relevant glucose concentrations (1-30 mM). One protein with a response close to this glucose range was identified, the GGBP triple mutant E149C, A213S, L238S with an equilibrium dissociation constant of 0.5 mM. These results suggest that biosensors for direct glucose detection based on SPR or similar refractive detection methods, if miniaturized, have the potential for development as continuous glucose monitoring devices. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:653 / 660
页数:8
相关论文
共 33 条
[1]   Kinetics of dissolution of Concanavalin A dextran sols in response to glucose measured by surface plasmon resonance [J].
Ballerstadt, R ;
Schultz, JS .
SENSORS AND ACTUATORS B-CHEMICAL, 1998, 46 (01) :50-55
[2]  
Boos W, 1974, Antibiot Chemother (1971), V19, P21
[3]   Biosensor analysis of the interaction between immobilized human serum albumin and drug compounds for prediction of human serum albumin binding levels [J].
Frostell-Karlsson, Å ;
Remaeus, A ;
Roos, H ;
Andersson, K ;
Borg, P ;
Hämäläinen, M ;
Karlsson, R .
JOURNAL OF MEDICINAL CHEMISTRY, 2000, 43 (10) :1986-1992
[4]   STRUCTURAL MECHANISMS FOR DOMAIN MOVEMENTS IN PROTEINS [J].
GERSTEIN, M ;
LESK, AM ;
CHOTHIA, C .
BIOCHEMISTRY, 1994, 33 (22) :6739-6749
[5]   Using receptor conformational change to detect low molecular weight analytes by surface plasmon resonance [J].
Gestwicki, JE ;
Hsieh, HV ;
Pitner, JB .
ANALYTICAL CHEMISTRY, 2001, 73 (23) :5732-5737
[6]   OPTICAL-FIBER SENSOR-BASED ON SURFACE-PLASMON EXCITATION [J].
HOMOLA, J .
SENSORS AND ACTUATORS B-CHEMICAL, 1995, 29 (1-3) :401-405
[7]   Detection of electrochemical enzymatic reactions by surface plasmon resonance measurement [J].
Iwasaki, Y ;
Horiuchi, T ;
Niwa, O .
ANALYTICAL CHEMISTRY, 2001, 73 (07) :1595-1598
[8]   A FIBEROPTIC CHEMICAL SENSOR-BASED ON SURFACE-PLASMON RESONANCE [J].
JORGENSON, RC ;
YEE, SS .
SENSORS AND ACTUATORS B-CHEMICAL, 1993, 12 (03) :213-220
[9]   SURFACE-PLASMON RESONANCE DETECTION AND MULTISPOT SENSING FOR DIRECT MONITORING OF INTERACTIONS INVOLVING LOW-MOLECULAR-WEIGHT ANALYTES AND FOR DETERMINATION OF LOW AFFINITIES [J].
KARLSSON, R ;
STAHLBERG, R .
ANALYTICAL BIOCHEMISTRY, 1995, 228 (02) :274-280