Automated Counting of Cells in Cerebrospinal Fluid Using the CellDyn-4000 Haematology Analyser

被引:31
作者
Hoffmann, Johannes J. M. L. [1 ]
Janssen, Willy C. M. [1 ]
机构
[1] Catharina Hosp, Dept Clin Labs, POB 1350, NL-5602 ZA Eindhoven, Netherlands
关键词
Cerebrospinal fluid; Differential count; Erythrocytes; Haematology analyser; Leukocytes;
D O I
10.1515/CCLM.2002.204
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 [基础医学];
摘要
Counting of cells in cerebrospinal fluid is currently performed manually. Because of the inherent analytical and economical disadvantages, we attempted to introduce a fully automated method. Therefore, we validated the Abbott CellDyn-4000 haematology analyser for counting cells in cerebrospinal fluid. The analyser was used in its standard configuration with the simple precaution of a preceding blank sample. As for leukocyte counting the analyser yielded high precision (CV similar to 5% above the upper reference limit), good linearity, low limit of detection (2/mu 1) and excellent correlation (r > 0.99) with the counting chamber method. The differential leukocyte count was equally accurate and precise, even in the low concentration range. Performance of the erythrocyte count was impaired by its high limit of detection (6/nl) and it appeared satisfactory only for detecting blood admixture due to traumatic puncture. The specificity of the analyser is excellent, since it correctly classified non-viable leukocytes and excluded yeast cells from the leukocyte count in a patient with cryptococcal meningitis. We conclude that the CellDyn-4000 is well suited for quickly and reliably counting leukocytes in cerebrospinal fluid. Developing some software modifications might make the analyser useful also for performing erythrocyte counting in cerebrospinal fluid.
引用
收藏
页码:1168 / 1173
页数:6
相关论文
共 9 条
[1]
Automated counting of white and red blood cells in the cerebrospinal fluid [J].
Aune, MW ;
Sandberg, S .
CLINICAL AND LABORATORY HAEMATOLOGY, 2000, 22 (04) :203-210
[2]
COULTER-COUNTER IDENTIFIES CRYPTOCOCCUS-NEOFORMANS AS LEUKOCYTES - A CASE OF PSEUDOPLEOCYTOSIS [J].
BERGER, SA ;
WAITMAN, DA ;
BARUCH, R .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1992, 97 (05) :663-664
[3]
Evaluation of the Abbott CELL-DYN 4000 hematology analyzer [J].
Grimaldi, E ;
Scopacasa, F .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2000, 113 (04) :497-505
[4]
Simultaneous differentiation and quantitation of erythroblasts and white blood cells on a high throughput clinical haematology analyser [J].
Kim, YR ;
Yee, M ;
Metha, S ;
Chupp, V ;
Kendall, R ;
Scott, CS .
CLINICAL AND LABORATORY HAEMATOLOGY, 1998, 20 (01) :21-29
[5]
KLEINE TO, 1991, EUR J CLIN CHEM CLIN, V29, P705
[6]
Mentz F, 1998, HEMATOL CELL THER, V40, P183
[7]
ELECTRONIC COUNTING OF SPINAL-FLUID CELLS [J].
TALSTAD, I .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1984, 81 (04) :506-511
[8]
Van Acker JT, 2001, CLIN CHEM, V47, P556
[9]
Ziebig R, 2000, CLIN CHEM, V46, P242