Direct high-performance liquid chromatographic determination of the enantiomeric purity of levodopa and methyldopa:: comparison with pharmacopoeial polarimetric methods

被引:53
作者
Dolezalová, M [1 ]
Tkaczyková, M [1 ]
机构
[1] State Inst Drug Control, Prague 10041 10, Czech Republic
关键词
levodopa; methyldopa; enantiomeric purity; chiral liquid chromatography; teicoplanin stationary phase; chiral ligand exchange phase; optical rotation;
D O I
10.1016/S0731-7085(98)00257-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Chiral high-performance liquid chromatography was employed for determination of the enantiomeric purity of levodopa and methyldopa. The determination of D-DOPA in levodopa was accomplished using a chiral ligand-exchange chromatograpy with an ordinary C-18 column and a chiral mobile phase containing N,N-dimethyl-L-phenylalanine and Cu(II) acetate or by means of LC on a teicoplanin column in conjunction with ethanol-water (65:35, v/v). Both methods gave good performance, however, the latter was faster and more convenient and suitable for routine analyses. For the determination of D-methyldopa a LC method based on the use of a teicoplanin column in polar organic mode with methanol-acetic acid-triethylamine (1000:0.05:0.05, v/v/v) mobile phase was developed. The precision, accuracy, linearity and selectivity were satisfactory. In comparison with pharmacopoeial polarimetric methods (according to the European Pharmacopoeia and the Pharmacopoea Bohemoslovaca), the LC methods proved to be much more sensitive giving detection limits 0.04% of D-DOPA and 0.3% of D-methyldopa. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:555 / 567
页数:13
相关论文
共 14 条
[1]  
*ADV SEP TECHN, 1996, CHIR HDB
[2]   COVALENTLY BONDED TEICOPLANIN CHIRAL STATIONARY-PHASE FOR HPLC ENANTIOSEPARATIONS [J].
ARMSTRONG, DW ;
LIU, YB ;
EKBORGOTT, KH .
CHIRALITY, 1995, 7 (06) :474-497
[3]   DIRECT LIQUID-CHROMATOGRAPHIC SEPARATION OF RACEMATES WITH AN ALPHA-CYCLODEXTRIN BONDED PHASE [J].
ARMSTRONG, DW ;
YANG, XF ;
HAN, SM ;
MENGES, RA .
ANALYTICAL CHEMISTRY, 1987, 59 (21) :2594-2596
[4]   Facile liquid chromatographic enantioresolution of native amino acids and peptides using a teicoplanin chiral stationary phase [J].
Berthod, A ;
Liu, YB ;
Bagwill, C ;
Armstrong, DW .
JOURNAL OF CHROMATOGRAPHY A, 1996, 731 (1-2) :123-137
[5]  
*COUNC EUR, 1996, EUR PAHRM
[6]  
*DAIC CHEM IND, APPL GUID CHIR COL S
[7]  
DAVANKOV VA, 1983, ADV CHROMATOGR, V22, P71
[8]  
DOYLE TD, 1991, ACS SYM SER, V471, P27
[9]   DETERMINATION OF THE ENANTIOMERIC PURITY OF LEVODOPA, METHYLDOPA, CARBIDOPA AND TRYPTOPHAN BY USE OF CHIRAL MOBILE PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
GELBER, LR ;
NEUMEYER, JL .
JOURNAL OF CHROMATOGRAPHY, 1983, 257 (02) :317-326
[10]   EVALUATION OF A CHIRAL CROWN-ETHER LC COLUMN FOR THE SEPARATION OF RACEMIC AMINES [J].
HILTON, M ;
ARMSTRONG, DW .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1991, 14 (01) :9-28