Wear of retrieved ceramic THA components - Four matched pairs retrieved after 5-13 years in service

被引:15
作者
Magnissalis, EA
Xenakis, TA
Zacharis, C
机构
[1] Res Ctr Biomat, Athens 16562, Greece
[2] Univ Ioannina, Orthopaed Clin, GR-45110 Ioannina, Greece
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 2001年 / 58卷 / 05期
关键词
in vivo wear; alumina; retrieval analysis; ceramic implants; total hip arthroplasty;
D O I
10.1002/jbm.1057.abs
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Retrieval analyses of alumina-alumina THA components are conducted in order to investigate in vivo wear of implants and provide guidance to material and design improvements. The contribution of the present study consists of the examination, by digital recording of spherical profiles and scanning electron microscopy, of four matched pairs of ceramic THA implants retrieved after 5, 11, 13, and 13 years in service because of mechanical loosening. Maximum wear depth values on measured surfaces ranged between 10-325 mum for heads and 25-784 mum for cups, while grossly worn areas varied between 240 and 1510 mm(2) for heads and 250 and 1570 mm(2) for cups. The greatest wear characteristics were demonstrated by the pair that served for 5 years, that is, on the cup that was found in a 65 degrees inclination. Quantitative topographical study of the results provided concrete evidence of a cascade of detrimental wear events of changing geometry, intensity, and progression associated with gradual cup tilting. Microscopy findings support previously recognized wear mechanisms. (C) 2001 John Wiley & Sons, Inc.
引用
收藏
页码:593 / 598
页数:6
相关论文
共 14 条
[1]
THE USE OF DENSE ALUMINA ALUMINA CERAMIC COMBINATION IN TOTAL HIP-REPLACEMENT [J].
BOUTIN, P ;
CHRISTEL, P ;
DORLOT, JM ;
MEUNIER, A ;
DEROQUANCOURT, A ;
BLANQUAERT, D ;
HERMAN, S ;
SEDEL, L ;
WITVOET, J .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (12) :1203-1232
[2]
CHRISTEL PS, 1992, CLIN ORTHOP RELAT R, P10
[3]
CLARKE IC, 1992, CLIN ORTHOP RELAT R, P19
[4]
DORLOT JM, 1992, CLIN ORTHOP RELAT R, P47
[5]
WEAR ANALYSIS OF RETRIEVED ALUMINA HEADS AND SOCKETS OF HIP PROSTHESES [J].
DORLOT, JM ;
CHRISTEL, P ;
MEUNIER, A .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH-APPLIED BIOMATERIALS, 1989, 23 (A3) :299-310
[6]
HARRIS WH, 1995, CLIN ORTHOP RELAT R, P46
[7]
*ISO, 1994, 6474 ISO
[8]
Lerouge S, 1996, J BIOMED MATER RES, V32, P627, DOI 10.1002/(SICI)1097-4636(199612)32:4<627::AID-JBM16>3.0.CO
[9]
2-A
[10]
MITTELMEIER H, 1992, CLIN ORTHOP RELAT R, P64