LC determination of morphine and morphine glucuronides in human plasma by coulometric and UV detection

被引:38
作者
Ary, K [1 ]
Róna, K [1 ]
机构
[1] Semmelweis Univ, Fac Hlth Sci, Dept Med 1, Clin Pharmacol Lab, H-1389 Budapest, Hungary
关键词
morphine; morphine-3-glucuronide; morphine-6-glucuronide; HPLC; simultaneous UV and ECD detection; solid-phase extraction;
D O I
10.1016/S0731-7085(01)00393-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A reversed-phase high-performance liquid chromatographic method with coulometric and UV detection has been developed for the simultaneous determination of morphine, morphine-3-glucuronide and morphine-6-glucuronide. The separation was carried out by using a Supelcosil LC-8 DB reversed-phase column and 0.1 M potassium dihydrogen phosphate (pH 2.5)-acetonitrile-methanol (94:5:1 v/v) containing 4 mM pentanesulfonic acid as the mobile phase. The compounds were determined simultaneously by coulometry for morphine and with UV detection for morphine-3-glucuronide and morphine-6-glucuronide. Morphine, morphine glucuronides and the internal standard were extracted from human plasma using Bond-Elut C18 (1 ml) solid-phase extraction cartridges. In the case of coulometric detection, the detection limit was 0.5 ng/ml for morphine; in the case of UV detection the detection limit was 10 ng/ml for morphine-3-glucuronide and for morphine-6-glucuronide. too. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:179 / 187
页数:9
相关论文
共 15 条