Retrievability of uncoated versus paclitaxel-coated Gunther-Tulip IVC filters in an animal model

被引:14
作者
De Gregorio, MA
Gimeno, MJ
Lostalé, F
Iñigo, P
Artigas, MC
Viloria, A
Alfonso, ER
D'Agostino, H
机构
[1] Univ Zaragoza, Intervent Radiol Sect, Inst Minimally Invas Tech, Zaragoza, Spain
[2] Louisiana State Univ, Hlth Sci Ctr, Dept Radiol, Shreveport, LA 71105 USA
关键词
D O I
10.1097/01.RVI.0000133512.20764.D8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To compare in a pilot study, the retrievability and inferior vena cava (IVC) wall reaction elicited by uncoated and paclitaxel-coated Gunther-Tulip filters in the animal model. MATERIALS AND METHODS: Three groups with five pigs each underwent infrarenal IVC implantation of Gunther-Tulip filters. Paclitaxel-coated filters were used in Groups A and B and uncoated filters were used in Group C. Filters were removed at 14, 19, 22, 26, and 30 days after implantation. A laparotomy was performed to remove filters from animals in group A and filters from animals in groups B and C that could not be retrieved via the right transjugular approach. Filter-induced venous wall changes were evaluated by examination of IVC venography, feasibility of filter removal at different implantation times, and laparotomy and microscopic findings. Feasibility of filter retrieval and venous wall changes were correlated. RESULTS: IVC cavography showed no abnormality. Filters in animals in group B were uneventfully removed by a right jugular approach. Uncoated filter removal was not feasible in three of five animals in group C (19, 22, and 26 days). Microscopically, animals in group A had absent filter-induced IVC wall changes at 14, 19, 22, and 26 days and minimal changes at 30 days post implantation; animals in group B had absent filter-induced IVC wall changes at 14, 19, and 22 days and minimal changes at 26 and 30 days; animals in group C had moderate filter-induced IVC wall changes at 14 days and severe changes at 19, 22, 26, and 30 days. CONCLUSIONS: This pilot study suggests that endothelial reaction to the presence of IVC filters in the porcine model is diminished by addition of paclitaxel coating to these filters. Further studies are necessary to substantiate these results.
引用
收藏
页码:719 / 726
页数:8
相关论文
共 26 条
[1]  
ABBOTT GT, 1999, J VASC INTERV RADIOL, V10, P288
[2]   Initial experience in humans with a new retrievable inferior vena cava filter [J].
Asch, MR .
RADIOLOGY, 2002, 225 (03) :835-844
[3]   Interactions of transforming growth factor-β and angiotensin II in renal fibrosis [J].
Border, WA ;
Noble, NA .
HYPERTENSION, 1998, 31 (01) :181-188
[4]  
BORDER WA, 1994, NEW ENGL J MED, V331, P1286
[5]   The Gunther Tulip retrievable filter:: Prolonged temporary filtration by repositioning within the inferior vena cava [J].
de Gregorio, MA ;
Gamboa, P ;
Gimeno, MJ ;
Madariaga, B ;
Tobío, R ;
Herrera, M ;
Medrano, J ;
Mainar, A ;
Alfonso, R .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (10) :1259-1265
[6]   Animal experience in the Gunther Tulip retrievable inferior vena cava filter [J].
de Gregorio, MA ;
Gimeno, MJ ;
Tobio, R ;
Lostalé, F ;
Mainar, A ;
Beltrán, JM ;
Madariaga, B ;
Alfonso, ER ;
Medrano, J ;
Viloria, A .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 24 (06) :413-417
[7]   A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis [J].
Decousus, H ;
Leizorovicz, A ;
Parent, F ;
Page, Y ;
Tardy, B ;
Girard, P ;
Laporte, S ;
Faivre, R ;
Charbonnier, B ;
Barral, FG ;
Huet, Y ;
Simonneau, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 338 (07) :409-415
[8]   EXPERIENCE WITH THE AMPLATZ RETRIEVABLE VENA-CAVA FILTER [J].
EPSTEIN, DH ;
DARCY, MD ;
HUNTER, DW ;
COLEMAN, CC ;
TADAVARTHY, SM ;
MURRAY, PD ;
CASTANEDAZUNIGA, WR ;
AMPLATZ, K .
RADIOLOGY, 1989, 172 (01) :105-110
[9]   Drug-eluting stents in vascular intervention [J].
Fattori, R ;
Piva, T .
LANCET, 2003, 361 (9353) :247-249
[10]  
Given MF, 2002, RADIOLOGY, V225, P642