Evaluation of four rapid methods for hemoglobin screening of whole blood donors in mobile collection settings

被引:41
作者
Gomez-Simon, Antonia [1 ]
Navarro-Nunez, Leyre [1 ]
Perez-Ceballos, Elena [1 ]
Lozano, Maria L. [1 ]
Candela, Maria J. [1 ]
Cascales, Almudena [1 ]
Martinez, Constantino [1 ]
Corral, Javier [1 ]
Vicente, Vicente [1 ]
Rivera, Jose [1 ]
机构
[1] Univ Murcia, Unit Hematol & Clin Oncol, Ctr Reg Hemodonac, Murcia 30003, Spain
关键词
blood donor; hemoglobin;
D O I
10.1016/j.transci.2007.01.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Predonation hemoglobin measurement is a problematic requirement in mobile donation settings, where accurate determination of venous hemoglobin by hematology analyzers is not available. Objective: We have evaluated hemoglobin screening in prospective donors by the serniquantitative copper sulphate test and by capillary blood samples analyzed by three portable photometers, HemoCue, STAT-Site((R)) MHgb, and the Compo-Lab HB system. Methods: Capillary blood samples were obtained from 380 donors and tested by the copper sulphate test and by at least one of the named portable photometers. Predonation venous hemoglobin was also determined in all donors using a Coulter Max-M analyzer. Results: The three photometers provided acceptable reproducibility (CV below 5%), and displayed a significant correlation between the capillary blood samples and the venous hemoglobin (R-2 0.5-0.8). HemoCue showed the best agreement with venous hemoglobin determination, followed by STAT-Site((R)) MHgb, and the Compo-Lab HB system. The copper sulphate test provided the highest rate of donors acceptance (83%) despite unacceptable hemoglobin levels, and the lowest rate for donor deferral (1%) despite acceptable hemoglobin levels. The percentage of donors correctly categorized for blood donation by the portable hemoglobinometers was 85%, 82%, and 76% for CompoLab HB system, HemoCue and STAT-Site respectively. Conclusion: Our data suggest that hemoglobin determination remains a conflictive issue in donor selection in the mobile setting. Without appropriate performance control, capillary hemoglobin screening by either the copper sulphate method or by the novel portable hemoglobinometers could be inaccurate, thus potentially affecting both donor safety and the blood supply. (c) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:235 / 242
页数:8
相关论文
共 19 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Managing donors and iron deficiency [J].
Boulton, F .
VOX SANGUINIS, 2004, 87 :22-24
[3]   IMPROVED STRATEGY FOR SCREENING PROSPECTIVE BLOOD-DONORS FOR ANEMIA [J].
BOULTON, FE ;
NIGHTINGALE, MJ ;
REYNOLDS, W .
TRANSFUSION MEDICINE, 1994, 4 (03) :221-225
[4]   A NEW VIEW OF INTER-OBSERVER AGREEMENT [J].
BURDOCK, EI ;
FLEISS, JL ;
HARDESTY, AS .
PERSONNEL PSYCHOLOGY, 1963, 16 (04) :373-384
[5]   Hb screening of blood donors: how close is close enough? [J].
Cable, RG .
TRANSFUSION, 2003, 43 (03) :306-308
[6]   HEMOGLOBIN DETERMINATION IN BLOOD-DONORS [J].
CABLE, RG .
TRANSFUSION MEDICINE REVIEWS, 1995, 9 (02) :131-144
[7]   EVALUATION OF METHODS AND PROTOCOLS FOR HEMOGLOBIN SCREENING OF PROSPECTIVE WHOLE-BLOOD DONORS [J].
CHAMBERS, LA ;
MCGUFF, JM .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1989, 91 (03) :309-312
[8]  
*COUNC EUR, 2004, GUID PREP US QUAL AS
[9]  
Fleiss J., 1986, Reliability of measurement: the design and analysis of clinical experiments
[10]   Statistical analysis of inappropriate results from current Hb screening methods for blood donors [J].
James, V ;
Jones, KE ;
Turner, EM ;
Sokol, RJ .
TRANSFUSION, 2003, 43 (03) :400-404