A simple and generally applicable method to estimate the peripheral dose in radiation teletherapy with high energy X-rays or gamma radiation

被引:48
作者
VanderGiessen, PH
机构
[1] Dr. Bernard Verbeeten Institute for Radiation Oncology and Nuclear Medicine, Department of Clinical Physics, Tilburg
[2] Dr. Bernard Verbeeten Institution for Radiation Oncology and Nuclear Medicine, 5042 SB , Tilburg
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1996年 / 35卷 / 05期
关键词
gonadal dose; peripheral dose; radiation therapy; risk assessment;
D O I
10.1016/0360-3016(96)00254-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Many articles have been published on the measurement of the dose to points outside the primary beam, often called the peripheral dose (PD), for instance, to the gonads, for specific treatment machines and/or techniques. We investigated the possibilities for developing a generalized method based on the data from several publications. Methods and Materials: The data from several publications were recalculated for a reference situation, then averaged, and the frequency distributions around the mean were determined. Published data were available for Co-60, 4, 6, 8, and 10 MV, and 18 to 25 MV for a large variety of treatment machines. Furthermore, an analysis of possible corrections for depth dependence, held elongation, irregularly shaped fields, wedges, and shielding blocks was carried out. Results: The frequency distributions of all published PD values for square fields for photon energies of 4 MV to 25 MV showed a standard deviation of 33%. The PD values of Co-60 are significantly different with a standard deviation of 25%. A difference in the leakage radiation between cobalt machines and linear accelerators can possibly explain this difference, especially for large distances, where leakage radiation predominates. Taking the uncertainty of the risk factors into consideration, we conclude that the use of average values is justified. Although statistically not significant, the peripheral dose appears to be dependent on photon energy with a minimum around 6 MV. Conclusions: It is possible to estimate the peripheral dose for photon energies of 4 MV to 25 MV with an accuracy of +/-33%; for Co-60, the accuracy is even better. The variation of the PD between different treatment machines is so small that it is justified to use average PD values, irrespective of the treatment machine.
引用
收藏
页码:1059 / 1068
页数:10
相关论文
共 26 条
[1]   PERIPHERAL DOSE FROM MEGAVOLT BEAMS [J].
FRAASS, BA ;
VANDEGEIJN, J .
MEDICAL PHYSICS, 1983, 10 (06) :809-818
[2]   CALCULATION OF THE DOSE DELIVERED TO ORGANS OUTSIDE THE RADIATION BEAMS [J].
FRANCOIS, P ;
BEURTHERET, C ;
DUTREIX, A .
MEDICAL PHYSICS, 1988, 15 (06) :879-883
[3]   DOSE LEVELS OUTSIDE RADIOTHERAPY BEAMS [J].
GREENE, D ;
KARUP, PGG ;
SIMS, C ;
TAYLOR, RJ .
BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (689) :453-456
[4]   DOSE LEVELS OUTSIDE RADIOTHERAPY BEAMS [J].
GREENE, D ;
CHU, GL ;
THOMAS, DW .
BRITISH JOURNAL OF RADIOLOGY, 1983, 56 (668) :543-550
[5]  
JENSEN JM, 1983, STRAHLENTHER ONKOL, V159, P555
[6]  
JOHNS HE, 1978, PHYSICS RADIOLOGY, P172
[7]  
KARTHA P, 1978, MED PHYS, V5, P327
[8]   MEASUREMENTS OF DOSE FROM SECONDARY RADIATION OUTSIDE A TREATMENT FIELD [J].
KASE, KR ;
SVENSSON, GK ;
WOLBARST, AB ;
MARKS, MA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (08) :1177-1183
[9]   A METHOD OF CALCULATING PERIPHERAL DOSE DISTRIBUTIONS OF PHOTON BEAMS BELOW 10-MV [J].
MCPARLAND, BJ ;
FAIR, HI .
MEDICAL PHYSICS, 1992, 19 (02) :283-293
[10]   SUPERHEATED DROP DETECTOR FOR DETERMINATION OF NEUTRON DOSE-EQUIVALENT TO PATIENTS UNDERGOING HIGH-ENERGY X-RAY AND ELECTRON RADIOTHERAPY [J].
NATH, R ;
MEIGOONI, AS ;
KING, CR ;
SMOLEN, S ;
DERRICO, F .
MEDICAL PHYSICS, 1993, 20 (03) :781-787