Monte Carlo simulation using DETECT2000 of a multilayered scintillation block and fit to experimental data

被引:11
作者
Cayouette, F [1 ]
Zhang, N
Thompson, CJ
机构
[1] McGill Univ, Dept Biochem Engn, Montreal, PQ H3A 2B4, Canada
[2] McGill Univ, Montreal Neurol Inst, Montreal, PQ H3A 2B4, Canada
[3] McGill Univ, Dept Biomed Engn, Montreal, PQ H3A 2B4, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
light transport; Monte Carlo simulation; pixellated crystals; positron emission mammography; surface finish;
D O I
10.1109/TNS.2003.812447
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The authors have used DETECT2000, a Monte Carlo simulation program of the light photon transport inside a scintillation block, to improve the performance of their next positron emission mammography system. The scintillation blocks of the system are composed of two arrays of crystals that are held together by an interconnecting layer. First, they used previously acquired experimental data on the block to find good simulated surface parameters for the scintillation block. They have found that the surface of the crystals are smooth and have good reflectivity. After a set of parameters was found to replicate the behavior of the block, the model geometry was modified in order to increase the block's performance. The DETECT2000 simulations determined that the removal of the interconnecting layer would yield the best results. The experimental results on a modified block yielded excellent results, showing the power of Monte Carlo simulations.
引用
收藏
页码:339 / 343
页数:5
相关论文
共 7 条
[1]   Monte Carlo modeling of scintillator crystal performance for stratified PET detectors with DETECT2000 [J].
Cayouette, F ;
Moisan, C ;
Zhang, N ;
Thompson, CJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (03) :624-628
[2]  
CAYOUETTE F, 2002, THESIS MCGILL U MONT
[3]  
HOSKINSON E, 1996, TRIDN9628 TRIUMF
[4]   LIGHT COLLECTION IN SCINTILLATION DETECTOR COMPOSITES FOR NEUTRON DETECTION [J].
KNOLL, GF ;
KNOLL, TF ;
HENDERSON, TM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :872-875
[5]  
MOISAN C, 1997, TRIPP9714 TRIUMF
[6]  
Murthy K, 2000, J NUCL MED, V41, P1851
[7]   Anode position and last dynode timing circuits for dual-layer BGO scintillator with PS-PMT based modular PET detectors [J].
Zhang, N ;
Thompson, CJ ;
Togane, D ;
Cayouette, F ;
Nguyen, KQ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (05) :2203-2207