New insights and observations in three-dimensional echocardiographic visualization of ventricular septal defects - Experimental and clinical studies

被引:27
作者
Kardon, RE [1 ]
Cao, QL [1 ]
Masani, N [1 ]
Sugeng, L [1 ]
Supran, S [1 ]
Warner, KG [1 ]
Pandian, NG [1 ]
Marx, GR [1 ]
机构
[1] Tufts Univ, Sch Med, New England Med Ctr, Boston Floating Hosp Infants & Children, Boston, MA 02111 USA
关键词
echocardiography; heart septal defects; pediatrics;
D O I
10.1161/01.CIR.98.13.1307
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-The positions, sizes, and shapes of ventricular septal defects (VSDs) can be difficult to assess by 2-dimensional echocardiography (2DE). Volume-rendered 3-dimensional echocardiography (3DE) can provide unique views of VSDs from the left ventricular (LV) side, allowing complete assessment of their circumference and spatial orientations to other anatomic structures. Methods and Results-Seventeen experimentally created defects of various locations, sizes, and shapes were imaged and reconstructed in 9 explanted porcine hearts. From an en face projection, major and minor axis diameters of the defects were measured, and these data were compared with direct anatomic measurements. Optimal reconstructions of the VSDs were obtained in all heart specimens, accurately depicting their positions and shapes. The correlations between 3DE and anatomy for the VSD major and minor axis diameters were y = 1.0x + 0.3 (r = 0.88, P<0.001) and y= 1.0x - 1.4 (r =0.89, P<0.001), respectively. Good agreement between the 2 methods was demonstrated for all measurements. Our experience from the in vitro model was then applied to patient studies. Optimal LV en face reconstructions were obtained in 45 of 51 patients, permitting detailed assessment of the positions, sizes, and shapes of the VSDs. In the 25 patients with comparative surgical measurements, the correlations between 3DE and surgery for the VSD major and minor axis diameters were y =0.81x + 2.1 (r=0.92, P<0.001) and y=0.73x + 2.0 (r=0.91, P<0.001), respectively. Good agreement was demonstrated between measurements made by 3DE and those obtained at surgery. Conclusions-3DE provides excellent visualization of various types of VSDs. From an LV en face projection, the positions, sizes, and shapes of VSDs can be accurately determined. Such precise imaging will be beneficial for surgical and catheter-based closure of difficult perimembranous and singular or multiple muscular VSDs.
引用
收藏
页码:1307 / 1314
页数:8
相关论文
共 20 条
[1]  
BAKER EJ, 1988, BRIT HEART J, V59, P339
[2]  
BIERMAN FZ, 1980, CIRCULATION, V62, P807, DOI 10.1161/01.CIR.62.4.807
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   PREOPERATIVE TRANSCATHETER CLOSURE OF CONGENITAL MUSCULAR VENTRICULAR SEPTAL-DEFECTS [J].
BRIDGES, ND ;
PERRY, SB ;
KEANE, JF ;
GOLDSTEIN, SAN ;
MANDELL, V ;
MAYER, JE ;
JONAS, RA ;
CASTENEDA, AR ;
LOCK, JE .
NEW ENGLAND JOURNAL OF MEDICINE, 1991, 324 (19) :1312-1317
[5]   CLASSIFICATION OF THE SITE OF VENTRICULAR SEPTAL-DEFECT BY TWO-DIMENSIONAL ECHOCARDIOGRAPHY [J].
CAPELLI, H ;
ANDRADE, JL ;
SOMERVILLE, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1983, 51 (09) :1474-1480
[6]   VENTRICULAR SEPTAL-DEFECT IN INFANCY - DETECTION WITH 2 DIMENSIONAL ECHOCARDIOGRAPHY [J].
CHEATHAM, JP ;
LATSON, LA ;
GUTGESELL, HP .
AMERICAN JOURNAL OF CARDIOLOGY, 1981, 47 (01) :85-89
[7]   MUSCULAR VENTRICULAR SEPTAL-DEFECTS REPAIRED WITH LEFT VENTRICULOTOMY [J].
GRIFFITHS, SP ;
TURI, GK ;
ELLIS, K ;
KRONGRAD, E ;
SWIFT, LH ;
GERSONY, WM ;
BOWMAN, FO ;
MALM, JR .
AMERICAN JOURNAL OF CARDIOLOGY, 1981, 48 (05) :877-886
[8]   STANDARDIZED NOMENCLATURE OF THE VENTRICULAR SEPTUM AND VENTRICULAR SEPTAL-DEFECTS, WITH APPLICATIONS FOR TWO-DIMENSIONAL ECHOCARDIOGRAPHY [J].
HAGLER, DJ ;
EDWARDS, WD ;
TAJIK, AJ ;
SEWARD, JB .
MAYO CLINIC PROCEEDINGS, 1985, 60 (11) :741-752
[9]  
KIRKLIN JK, 1980, J THORAC CARDIOV SUR, V80, P485
[10]   Volume-rendered, three-dimensional echocardiographic determination of the size, shape, and position of atrial septal defects: Validation in an in vitro model [J].
Magni, G ;
Cao, QL ;
Sugeng, L ;
Delabays, A ;
Marx, G ;
Ludomirski, A ;
Vogel, M ;
Pandian, NG .
AMERICAN HEART JOURNAL, 1996, 132 (02) :376-381