Artifact-resistant power-efficient design of finger-ring plethysmographic sensors

被引:154
作者
Rhee, S [1 ]
Yang, BH [1 ]
Asada, HH [1 ]
机构
[1] MIT, Dept Mech Engn, Arberloff Lab Informat Syst & Technol, Cambridge, MA 02139 USA
关键词
ambulatory monitoring; beat-to-beat pulsation; motion artifact; plethysmograph; power consumption; ring sensor; telemetry; wearable sensor;
D O I
10.1109/10.930904
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A miniaturized, telemetric, photoplethysmograph (PPG) sensor for long-term, continuous monitoring is presented in this paper. The sensor, called a "ring sensor," is attached to a finger base for monitoring beat-to-beat pulsation, and the data is sent to a host computer via a radio-frequency transmitter. Two major design issues are addressed: one is to minimize motion artifact and the other is to minimize the consumption of battery power. An efficient double ring design is developed to lower the influence of external force, acceleration, and ambient light, and to hold the sensor gently and securely on the skin, so that the circulation at the finger may not be obstructed. Total power consumption is analyzed in relation to characteristics: of individual components, sampling rate, and CPU clock speed. Optimal operating conditions are obtained for minimizing the power budget. A prototype ring sensor is designed and built based on the power budget analysis and the artifact-resistive attachment method. It is verified through experiments that the ring sensor is resistant to interfering forces and acceleration acting on the ring body. Benchmarking tests with FDA-approved PPG and electrocardiogram reveal that the ring sensor is comparable to those devices in detecting beat-to-beat pulsation despite disturbances.
引用
收藏
页码:795 / 805
页数:11
相关论文
共 12 条
[1]   CONTINUOUS ELECTROCARDIOGRAPHIC MONITORING DURING AUTOMOBILE DRIVING - STUDIES IN NORMAL SUBJECTS AND PATIENTS WITH CORONARY DISEASE [J].
BELLET, S ;
ROMAN, L ;
KOSTIS, J ;
SLATER, A .
AMERICAN JOURNAL OF CARDIOLOGY, 1968, 22 (06) :856-&
[2]  
DANIEL RK, 1981, ARCH PHYS MED REHAB, V62, P492
[3]  
DAVID D, 1988, AMBULATORY MONITORIN, pCH1
[4]  
DINSDALE SM, 1970, THESIS U MINNESOTA
[5]  
HAYES MJ, 1998, P CNVD 97 FRONTIERS, P117
[7]  
KOSIAK M, 1961, Arch Phys Med Rehabil, V42, P19
[8]   MEASUREMENT OF PULSE-WAVE VELOCITY USING A BEAT-SAMPLING TECHNIQUE [J].
PRUETT, JD ;
BOURLAND, JD ;
GEDDES, LA .
ANNALS OF BIOMEDICAL ENGINEERING, 1988, 16 (04) :341-347
[9]   Ischemic and pressure-induced hyperemia: A comparison [J].
Rendell, MS ;
Wells, JM .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1998, 79 (11) :1451-1455
[10]  
RHEE S, 1998, 20 ANN INT C IEEE EN