Evaluation of systematic quantification errors in velocity-selective arterial spin labeling of the brain

被引:59
作者
Duhamel, G
de Bazelaire, U
Alsop, DC
机构
[1] Beth Israel Deaconess Med Ctr, Dept Radiol, Ctr Adv Imaging W CC090, Boston, MA 02215 USA
[2] Harvard Univ, Sch Med, Boston, MA USA
关键词
arterial spin labeling; perfusion imaging; blood flow; stroke;
D O I
10.1002/mrm.10510
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Velocity-selective (VS) sequences potentially permit arterial spin labeling (ASL) perfusion imaging with labeling applied very close to the tissue. In this study the effects of cerebrospinal fluid (CSF) motion, radiofrequency (RF) field imperfections, and sequence timing parameters on the appearance and quantitative perfusion values obtained with VS-ASL were evaluated. Large artifacts related to CSF motion were observed with moderate velocity weighting, which were removed by inversion recovery preparation at the cost of increased imaging time. Imperfect refocusing and excitation pulses resulting from nonuniform RF fields produced systematic errors in the ASL subtraction images. A phase cycling scheme was introduced to eliminate these errors. Quantitative perfusion images were obtained with CSF suppression and phase cycling. Gray matter blood flow of 27.7 ml 100 g(-1) min(-1), approximately half the value reported in studies using spatially-selective ASL, was measured. Potential sources for this underestimation are discussed. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:145 / 153
页数:9
相关论文
共 22 条
[1]   THE EFFECTS OF RANDOM DIRECTIONAL DISTRIBUTED FLOW IN NUCLEAR-MAGNETIC-RESONANCE IMAGING [J].
AHN, CB ;
LEE, SY ;
NALCIOGLU, O ;
CHO, ZH .
MEDICAL PHYSICS, 1987, 14 (01) :43-48
[2]   Radio frequency magnetic field mapping of a 3 tesla birdcage coil: Experimental and theoretical dependence on sample properties [J].
Alecci, M ;
Collins, CM ;
Smith, MB ;
Jezzard, P .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (02) :379-385
[3]   Reduced transit-time sensitivity in noninvasive magnetic resonance imaging of human cerebral blood flow [J].
Alsop, DC ;
Detre, JA .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1996, 16 (06) :1236-1249
[4]   A general kinetic model for quantitative perfusion imaging with arterial spin labeling [J].
Buxton, RB ;
Frank, LR ;
Wong, EC ;
Siewert, B ;
Warach, S ;
Edelman, RR .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (03) :383-396
[5]   Magnetic resonance perfusion imaging in acute ischemic stroke using continuous arterial spin labeling [J].
Chalela, JA ;
Alsop, DC ;
Gonzalez-Atavales, JB ;
Maldjian, JA ;
Kasner, SE ;
Detre, JA .
STROKE, 2000, 31 (03) :680-687
[6]   Noninvasive MRI evaluation of cerebral blood flow in cerebrovascular disease [J].
Detre, JA ;
Alsop, DC ;
Vives, LR ;
Maccotta, L ;
Teener, JW ;
Raps, EC .
NEUROLOGY, 1998, 50 (03) :633-641
[7]   PERFUSION IMAGING [J].
DETRE, JA ;
LEIGH, JS ;
WILLIAMS, DS ;
KORETSKY, AP .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (01) :37-45
[8]   QUALITATIVE MAPPING OF CEREBRAL BLOOD-FLOW AND FUNCTIONAL LOCALIZATION WITH ECHO-PLANAR MR-IMAGING AND SIGNAL TARGETING WITH ALTERNATING RADIO-FREQUENCY [J].
EDELMAN, RR ;
SIEWERT, B ;
DARBY, DG ;
THANGARAJ, V ;
NOBRE, AC ;
MESULAM, MM ;
WARACH, S .
RADIOLOGY, 1994, 192 (02) :513-520
[9]   NORMAL FLOW PATTERNS OF INTRACRANIAL AND SPINAL CEREBROSPINAL-FLUID DEFINED WITH PHASE-CONTRAST CINE MR IMAGING [J].
ENZMANN, DR ;
PELC, NJ .
RADIOLOGY, 1991, 178 (02) :467-474
[10]   REAL-TIME FLOW MEASUREMENTS USING ECHO-PLANAR IMAGING [J].
GUILFOYLE, DN ;
GIBBS, P ;
ORDIDGE, RJ ;
MANSFIELD, P .
MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (01) :1-8