Micropower Non-contact EEG Electrode with Active Common-Mode Noise Suppression and Input Capacitance Cancellation

被引:56
作者
Chi, Yu M. [1 ]
Cauwenberghs, Gert [2 ]
机构
[1] Univ Calif San Diego, Dept Elect & Comp Engn, San Diego, CA 92092 USA
[2] Univ Calif San Diego, Jacobs Sch Engn, Div Biol Sci & Bioengn, San Diego, CA 92092 USA
来源
2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20 | 2009年
基金
美国国家科学基金会;
关键词
D O I
10.1109/IEMBS.2009.5333527
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A non-contact EEG electrode with input capacitance neutralization and common-mode noise suppression circuits is presented. The coin sized sensor capacitively couples to the scalp without direct contact to the skin. To minimize the effect of signal attenuation and channel gain mismatch, the input capacitance of each sensor is actively neutralized using positive feedback and bootstrapping. Common-mode suppression is achieved through a single conductive sheet to establish a common mode reference. Each sensor electrode provides a differential gain of 60dB. Signals are transmitted in a digital serial daisy-chain directly from a local 16-bit ADC, minimizing the number of wires required to establish a high density EEG sensor network. The micropower electrode consumes only 600 mu W from a single 3.3V supply.
引用
收藏
页码:4218 / +
页数:2
相关论文
共 15 条
[1]  
Aleksandrowicz A, 2007, ACTA POLYTECH, V47, P68
[2]  
CHI Y, 2009, IEEE BODY S IN PRESS
[3]  
DORNHEGE G, 2009, BRAIN COMPUTER INTER
[5]   Electric potential probes - new directions in the remote sensing of the human body [J].
Harland, CJ ;
Clark, TD ;
Prance, RJ .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2002, 13 (02) :163-169
[6]  
Larry, 1996, CAPACITIVE SENSORS D
[7]   CAPACITIVE ELECTROCARDIOGRAPHIC AND BIOELECTRIC ELECTRODES [J].
LOPEZ, A ;
RICHARDSON, PC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1969, BM16 (01) :99-+
[8]   ECG measurement using capacitive coupling electrodes for man-machine emotional communication [J].
Maruyama, T. ;
Makikawa, M. ;
Shiozawa, N. ;
Fujiwara, Y. .
2007 IEEE/ICME INTERNATIONAL CONFERENCE ON COMPLEX MEDICAL ENGINEERING, VOLS 1-4, 2007, :378-383
[9]   BARIUM-TITANATE CERAMICS CAPACITIVE-TYPE EEG ELECTRODE [J].
MATSUO, T ;
IINUMA, K ;
ESASHI, M .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (04) :299-300
[10]  
MATTHEWS R, 2007, NOVEL HYBRID BIOELEC