Mammography fixed grid versus reciprocating grid: Evaluation using cadaveric breasts as test objects

被引:9
作者
KimmeSmith, C
Sayre, J
McCombs, M
Gold, RH
Bassett, LW
机构
[1] Iris Cantor Ctr. for Breast Imaging, UCLA, School of Medicine, Los Angeles
[2] Radiological Sciences - CHS, UCLA, School of Medicine, Los Angeles
关键词
D O I
10.1118/1.597695
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this study we use unfixed cadaveric breasts to obtain mammography images with fixed and reciprocating grids. Sheets of acrylic, containing one or more clusters of simulated calcifications and masses, were superimposed on two fresh cadaveric breasts (3.4 and 6.5 cm thick), and were imaged with a fixed grid and a reciprocating grid. Six radiologists, working independently, attempted to identify the number of clusters and/or masses in 114 images containing 139 clusters of simulated calcifications and 42 simulated masses. Thirty-four of these images were normal, containing no lesions. For the thinner breast, no statistically significant difference was found in the detection of clusters of calcifications in the images produced with the fixed grid compared to those produced with the reciprocating grid. However, for the detection of calcifications in images of the thicker breast, sensitivity of 74% for detection of calcifications when a fixed grid was used was significantly less than sensitivity of 86% when a reciprocating grid was used (P = 0.006). The mass detection sensitivity was 91% for images made with a fixed grid compared to 96% for images made with a reciprocating grid, but the difference was not statistically significant (P = 0.346). The use of cadaveric breasts as test objects was well accepted by radiologists. Only for the thick cadaveric breast were differences between the two grids significant, and these differences were restricted to the task of finding calcifications. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:141 / 147
页数:7
相关论文
共 14 条
  • [2] CALCULATION OF DOSE AND CONTRAST FOR 2 MAMMOGRAPHIC GRIDS
    DANCE, DR
    PERSLIDEN, J
    CARLSSON, GA
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) : 235 - 248
  • [3] FEIG SA, 1987, RADIOL CLIN N AM, V25, P897
  • [4] DUAL-ENERGY MAMMOGRAPHY - INITIAL EXPERIMENTAL RESULTS
    JOHNS, PC
    DROST, DJ
    YAFFE, MJ
    FENSTER, A
    [J]. MEDICAL PHYSICS, 1985, 12 (03) : 297 - 304
  • [5] Kopans D.B., 1993, Breast Imaging Reporting and Data System, vA American College of
  • [6] LESION CONSPICUITY, STRUCTURED NOISE, AND FILM READER ERROR
    KUNDEL, HL
    REVESZ, G
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1976, 126 (06) : 1233 - 1238
  • [7] SCANNED-PROJECTION DIGITAL MAMMOGRAPHY
    NISHIKAWA, RM
    MAWDSLEY, GE
    FENSTER, A
    YAFFE, MJ
    [J]. MEDICAL PHYSICS, 1987, 14 (05) : 717 - 727
  • [8] COMPARISON OF 2 SCREEN-FILM COMBINATIONS IN CONTACT AND MAGNIFICATION MAMMOGRAPHY - DETECTABILITY OF MICROCALCIFICATIONS
    OESTMANN, JW
    KOPANS, DB
    LINETSKY, L
    HALL, DA
    MCCARTHY, KA
    WHITE, G
    SWANN, C
    KELLEY, JE
    JOHNSON, LL
    [J]. RADIOLOGY, 1988, 168 (03) : 657 - 659
  • [9] BREAST CALCIFICATIONS - ANALYSIS OF IMAGING PROPERTIES
    OLSON, SL
    FAM, BW
    WINTER, PF
    SCHOLZ, FJ
    LEE, AK
    GORDON, SE
    [J]. RADIOLOGY, 1988, 169 (02) : 329 - 332
  • [10] Potter HE, 1920, AM J ROENTGENOL, V7, P292