Reliability and validity of the combined heart rate and movement sensor Actiheart

被引:476
作者
Brage, S
Brage, N
Franks, PW
Ekelund, U
Wareham, NJ
机构
[1] Univ Cambridge, Inst Publ Hlth, MRC, Epidemiol Unit,Elsie Widdowson Lab, Cambridge CB1 9NL, England
[2] Univ So Denmark, Inst Sports Sci & Clin Biomech, Odense, Denmark
基金
英国惠康基金;
关键词
physical activity; heart rate variability; accelerometer; movement; walking; running; epidemiology; energy expenditure monitoring;
D O I
10.1038/sj.ejcn.1602118
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Accurate quantification of physical activity energy expenditure is a key part of the effort to understand disorders of energy metabolism. The Actiheart, a combined heart rate (HR) and movement sensor, is designed to assess physical activity in populations. Objective: To examine aspects of Actiheart reliability and validity in mechanical settings and during walking and running. Methods: In eight Actiheart units, technical reliability ( coefficients of variation, CV) and validity for movement were assessed with sinusoid accelerations (0.1-20 m/s(2)) and for HR by simulated R-wave impulses (25-250 bpm). Agreement between Actiheart and ECG was determined during rest and treadmill locomotion (3.2-12.1 km/h). Walking and running intensity (in J/min/kg) was assessed with indirect calorimetry in 11 men and nine women (26-50y, 20-29 kg/m(2)) and modelled from movement, HR, and movement+HR by multiple linear regression, adjusting for sex. Results: Median intrainstrument CV was 0.5 and 0.03% for movement and HR, respectively. Corresponding interinstrument CV values were 5.7 and 0.03% with some evidence of heteroscedasticity for movement. The linear relationship between movement and acceleration was strong (R-2=0.99, P<0.001). Simulated R-waves were detected within 1 bpm from 30 to 250 bpm. The 95% limits of agreement between Actiheart and ECG were -4.2 to 4.3 bpm. Correlations with intensity were generally high (R-2>40.84, P<0.001) but significantly highest when combining HR and movement (SEE<1 MET). Conclusions: The Actiheart is technically reliable and valid. Walking and running intensity may be estimated accurately but further studies are needed to assess validity in other activities and during free-living. Sponsorship: The study received financial support from the Wellcome Trust and SB was supported by a scholarship from Unilever, UK.
引用
收藏
页码:561 / 570
页数:10