Effect of synovial fluid, phosphate-buffered saline solution, and water on the dissolution and corrosion properties of CoCrMo alloys as used in orthopedic implants

被引:78
作者
Lewis, AC
Kilburn, MR
Papageorgiou, I
Allen, GC
Case, CP
机构
[1] Univ Bristol, Interface Anal Ctr, Bristol BS2 8BS, Avon, England
[2] Southmead Gen Hosp, Bristol Implant Res Ctr, Bristol BS10 5NB, Avon, England
关键词
cobalt alloy; corrosion; metal ion release; passivation; protein film; XPS;
D O I
10.1002/jbm.a.30368
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The corrosion and dissolution of high- and low-carbon CoCrMo alloys, as used in orthopedic joint replacements, were studied by immersing samples in phosphate-buffered saline (PBS), water, and synovial fluid at 37° C for up to 35 days. Bulk properties were analyzed with a fine ion beam microscope. Surface analyses by X-ray photoelectron spectroscopy and Auger electron spectroscopy showed surprisingly that synovial fluid produced a thin oxide/hydroxide layer. Release of ions into solution from the alloy also followed an unexpected pattern where synovial fluid, of all the samples, had the highest Cr concentration but the lowest Co concentration. The presence of carbide inclusions in the alloy did not affect the corrosion or the dissolution mechanisms, although the carbides were a significant feature on the metal surface. Only one mechanism was recognized as controlling the thickness of the oxide/hydroxide interface. The analysis of the dissolved metal showed two mechanisms at work: (1) a protein film caused ligand-induced dissolution, increasing the Cr concentration in synovial fluid, and was explained by the equilibrium constants; (2) corrosion at the interface increased the Co in PBS. The effect of prepassivating the samples (ASTM F-86-01) did not always have the desired effect of reducing dissolution. The release of Cr into PBS increased after prepassivation. The metal-synovial fluid interface did not contain calcium phosphate as a deposit, typically found where samples are exposed to calcium rich bodily fluids. © 2005 Wiley Periodicals, Inc.
引用
收藏
页码:456 / 467
页数:12
相关论文
共 34 条
[1]  
[Anonymous], F75600 ASTM
[2]  
[Anonymous], 1996, 5707 ISO, P1
[3]  
ASTM, 2001, Alcohol, V86, P1, DOI [10.1520/D1957-86R01, DOI 10.1520/D1957-86R01, DOI 10.1520/F2129-08]
[4]  
Bard A.J., 1985, MONOGRAPHS ELECTROAN
[5]  
Beamson G., 1992, ADV MATER, DOI DOI 10.1002/ADMA.19930051035
[6]   WIDESPREAD DISSEMINATION OF METAL DEBRIS FROM IMPLANTS [J].
CASE, CP ;
LANGKAMER, VG ;
JAMES, C ;
PALMER, MR ;
KEMP, AJ ;
HEAP, PF ;
SOLOMON, L .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (05) :701-712
[7]   Chromosomal changes after surgery for joint replacement [J].
Case, CP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (08) :1093-1095
[8]  
CASE CP, 1996, CLIN ORTHOP S, V329, P269
[9]  
Chan FW, 1996, CLIN ORTHOP RELAT R, P96
[10]  
DAY JCC, PROGRAMMABLE INTERFA