Bioelectrical impedance measurements in patients with gastrointestinal disease: validation of the spectrum approach and a comparison of different methods for screening for nutritional depletion

被引:33
作者
Cox-Reijven, PLM
van Kreel, B
Soeters, PB
机构
[1] Univ Hosp Maastricht, Dept Dietet, NL-6202 AZ Maastricht, Netherlands
[2] Univ Hosp Maastricht, Dept Clin Chem, NL-6202 AZ Maastricht, Netherlands
[3] Univ Hosp Maastricht, Dept Surg, NL-6202 AZ Maastricht, Netherlands
关键词
patients; gastrointestinal disease; depletion; body composition; bioelectrical impedance spectroscopy; BIS; total body water; extracellular water; intracellular water; screening;
D O I
10.1093/ajcn/78.6.1111
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 [营养与食品卫生学];
摘要
Background: Bioelectrical impedance spectroscopy (BIS) may be of value in identifying patients with nutritional depletion. Objective: The primary aim of the study was to assess the validity of BIS in patients with gastrointestinal disease. The secondary aim was to compare different screening methods for identifying nutritionally depleted patients. Design: In 70 patients with gastrointestinal disease, BIS measurements were performed and fluid compartments were calculated from extrapolated resistance (R) values of extracellular (R-ECW) and intracellular (R-ICW) water based on mixture equations. Results were compared with total body water and ECW, which were measured by dilution methods. Patients were classified as depleted if their actual fat-free mass (FFM) was < 90% of their ideal FFM (iFFM). Different screening approaches for depletion were compared: the impedance vector method and the resistance index (RI) in relation to iFFM (RI/iFFM). Results: Total body water and ICW were largely underestimated (4 L) in the not-depleted patients but not in the depleted patients. In both groups, ECW was slightly overestimated (0.6-0.7 L). The critical frequency was 60% higher and the membrane capacitance was 40% lower in the depleted than in the not-depleted patients. The impedance vector method failed to identify depleted patients, whereas the proposed ratio at R,, (R measured at 50 kHz) and R-inf (R at infinite frequency) resulted in comparable sensitivity (86%) and specificity (73-80%). Conclusions: BIS measures of body fluids are influenced by the presence of depletion, probably because of alterations in the electric properties of the body at the cellular level. However, for screening purposes, the use of the ratio (L-2/R)/iFFM at R-50 and R-inf may be of value.
引用
收藏
页码:1111 / 1119
页数:9
相关论文
共 36 条
[1]
[Anonymous], 1983, Stat Bull Metrop Life Found, V64, P3
[2]
Hydration status assessed by multi-frequency bioimpedance analysis [J].
Asselin, MC ;
Kriemler, S ;
Chettle, DR ;
Webber, CE ;
Bar-Or, O ;
McNeill, FE .
APPLIED RADIATION AND ISOTOPES, 1998, 49 (5-6) :495-497
[3]
Body-water compartments measured by bio-electrical impedance spectroscopy in patients with chronic obstructive pulmonary disease [J].
Baarends, EM ;
Lichtenbelt, WDV ;
Wouters, EFM ;
Schols, AMWJ .
CLINICAL NUTRITION, 1998, 17 (01) :15-22
[4]
Bioelectrical impedance analysis during acute changes of extracellular osmolality in man [J].
Berneis, K ;
Keller, U .
CLINICAL NUTRITION, 2000, 19 (05) :361-366
[5]
BERRY JK, 1998, CRIT CARE NURS Q, V21, P22
[6]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]
Role of bioimpedance spectroscopy in assessment of body water compartments in hemodialysis patients [J].
Cox-Reijven, PL ;
Kooman, JP ;
Soeters, PB ;
van der Sande, FM ;
Leunissen, KML .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2001, 38 (04) :832-838
[8]
Predicting body cell mass with bioimpedance by using theoretical methods: A technological review [J].
DeLorenzo, A ;
Andreoli, A ;
Matthie, J ;
Withers, P .
JOURNAL OF APPLIED PHYSIOLOGY, 1997, 82 (05) :1542-1558
[9]
Bioelectrical impedance methods in clinical research: A follow-up to the NIH technology assessment conference [J].
Ellis, KJ ;
Bell, SJ ;
Chertow, GM ;
Chumlea, WC ;
Knox, TA ;
Kotler, DP ;
Lukaski, HC ;
Schoeller, DA .
NUTRITION, 1999, 15 (11-12) :874-880
[10]
Progressive cellular dehydration and proteolysis in critically ill patients [J].
Finn, PJ ;
Plank, LD ;
Clark, MA ;
Connolly, AB ;
Hill, GL .
LANCET, 1996, 347 (9002) :654-656