Electrophoretic separation of recombinant tissue-type plasminogen activator glycoforms: Validation issues for capillary isoelectric focusing methods

被引:30
作者
Moorhouse, KG
Rickel, CA
Chen, AB
机构
[1] Dept. of Qual. Contr. Clin. Devmt., Genentech Inc., South San Francisco, CA
[2] Genentech Inc., So. San Francisco, CA 94080
关键词
capillary isoelectric focusing; recombinant tissue-type plasminogen activator; capillary electrophoresis; validation;
D O I
10.1002/elps.1150170223
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Attempts were made to validate a capillary isoelectric focusing (cIEF) method for a recombinant glycoprotein as an alternative technique to slab gel isoelectric focusing methods routinely used to monitor such charge heterogeneity. The cIEF method principally separates the charged glycoforms of recombinant tissue-type plasminogen activator (rt-PA) on the basis of their sialic acid content. Nine to ten distinct peaks were consistently resolved, with the profile dependent on the class of ampholyte used. The pI of rt-PA measured with synthetic pI standards was in the range pH 6.5-7.5 with the migration of the standards affected by the presence of the protein. The method showed an acceptable recovery of >100% and had good sensitivity where 25 ng of protein could be resolved into constituent peaks. Recovery of both major peaks and total protein measured by peak areas was linear over a wide range from 50-1000 mu g/mL. A detailed study showed that when a capillary had been used fbr some time, capillary age affected peak migration times and, to a lesser extent, resolution. Peak migration times were stable over a temperature range of 15-30 degrees C, and decreased predictably with increasing voltages (400-600 V/cm) and decreasing N, N, N', N'-tetramethylethylene diamine (TEMED) concentrations (0.4-1.5% v/v). Overall the data indicated that this methodology has the potential to be used in the commercial release of protein pharmaceuticals if variability resulting from capillary age and lot were resolved. Even in its present format the method equals the performance of slab gel IEF whilst offering significant improvements in ease of operation and in time and reagent use.
引用
收藏
页码:423 / 430
页数:8
相关论文
共 12 条
[1]   THE ROLE OF QUALITY-CONTROL IN BIOTECHNOLOGY - AN ANALYTICAL PERSPECTIVE [J].
GARNICK, RL ;
SOLLI, NJ ;
PAPA, PA .
ANALYTICAL CHEMISTRY, 1988, 60 (23) :2546-2557
[2]   ZONE BROADENING IN ELECTROPHORESIS WITH SPECIAL REFERENCE TO HIGH-PERFORMANCE ELECTROPHORESIS IN CAPILLARIES - AN INTERPLAY BETWEEN THEORY AND PRACTICE [J].
HJERTEN, S .
ELECTROPHORESIS, 1990, 11 (09) :665-690
[3]  
Mazzeo Jeff R., 1992, Methods (Orlando), V4, P205, DOI 10.1016/1046-2023(92)90035-7
[4]   RAPID ONE-STEP CAPILLARY ISOELECTRIC-FOCUSING METHOD TO MONITOR CHARGED GLYCOFORMS OF RECOMBINANT HUMAN TISSUE-TYPE PLASMINOGEN-ACTIVATOR [J].
MOORHOUSE, KG ;
EUSEBIO, CA ;
HUNT, G ;
CHEN, AB .
JOURNAL OF CHROMATOGRAPHY A, 1995, 717 (1-2) :61-69
[5]  
PRITCHETT TJ, 1994, 6 INT S HIGH PERF CA
[6]   APPLICATIONS OF CAPILLARY ELECTROPHORESIS IN PHARMACEUTICAL ANALYSIS [J].
RABEL, SR ;
STOBAUGH, JF .
PHARMACEUTICAL RESEARCH, 1993, 10 (02) :171-186
[7]  
SEIFRIED E, 1988, ARZNEIMITTEL-FORSCH, V38-1, P418
[8]  
SLAIS K, 1994, J CHROMATOGR A, V661, P249
[9]  
SPELLMAN MW, 1989, J BIOL CHEM, V264, P14100
[10]   USE OF HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS TO MONITOR CHARGE HETEROGENEITY IN RECOMBINANT-DNA DERIVED PROTEINS [J].
WU, SL ;
TESHIMA, G ;
CACIA, J ;
HANCOCK, WS .
JOURNAL OF CHROMATOGRAPHY, 1990, 516 (01) :115-122