Use of reversed-phase high-performance liquid chromatography-diode array detection for complete separation of 2,4,6-trinitrotoluene metabolites and EPA method 8330 explosives: influence of temperature and an ion-pair reagent

被引:53
作者
Borch, T [1 ]
Gerlach, R [1 ]
机构
[1] Montana State Univ, Ctr Biofilm Engn, Bozeman, MT 59717 USA
关键词
explosives; environmental analysis; EPA method 8330; temperature selectivity; 2,4,6-trinitrotoluene; hydroxylaminodinitrotoluene; tetranitroazoxytoluene; tetranitroazotoluene;
D O I
10.1016/j.chroma.2003.09.067
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Explosives such as 2,4,6-trinitrotoluene (TNT), octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX), and hexahydro-1,3,5-trinitro-1,3,5-triazine (RDX) are widely distributed environmental contaminants. Complete chromatographic separation is necessary in order to accurately determine and quantify explosives and their degradation products in environmental samples and in (bio)transformation studies. The present study describes a RP-HPLC method with diode array detection using a LC-8 guard column, a Supelcosil LC-8 chromatographic column, and a gradient elution system. This gradient method is capable of baseline separating the most commonly observed explosives and TNT transformation metabolites including 2,4,6-triaminotoluene (TAT) in a single run. The TNT metabolites separated were 2-hydroxylamino-4,6-dinitrotoluene, 4-hydroxylamino-2,6-dinitrotoluene, 2,4-dihydroxylamino-6-nitrotoluene, 4,4',6,6'-tetranitro-2,2-azoxytoluene, 2,2',6,6'-tetranitro-4,4'-azoxytoluene, 4,4'.6,6'-tetranitro-2,2'-azotoluene, 2,2',6,6-tetranitro-4,4'-azotoluene, 2-amino-4,6-dinitrotoluene, 4-amino-2, 6-dinitrotoluene, 2,6-diamino-4-nitrotoluene, 2,4-diamino-6-nitrotoluene, and TAT. The same gradient method at a different column temperature can also be used to baseline separate the explosives targeted in the Environmental Protection Agency (EPA) Method 8330 with approximately 22% reduction in total run time and 48% decrease in solvent consumption compared to previously published methods. Good separation was also obtained when all TNT metabolites and EPA Method 8330 compounds (a total of 23 compounds) were analyzed together; only 2,6-DANT and HMX co-eluted in this case. The influence of temperature (35-55 degreesC) and the use of an ion-pair reagent on the chromatographic resolution and retention were investigated. Temperature was identified as the key parameter for optimal baseline separation. Increased temperature resulted in shorter retention times and better peak resolution especially for the aminoaromatics investigated. The use of an ion-pair reagent (octanesulfonic acid) generally resulted in longer retention times for compounds containing amine functional groups, more baseline noise, and decreased peak resolution. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:83 / 94
页数:12
相关论文
共 38 条
[1]   USE OF NARROW-BORE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY DIODE-ARRAY DETECTION FOR THE ANALYSIS OF INTERMEDIATES OF THE BIOLOGICAL DEGRADATION OF 2,4,6-TRINITROTOLUENE [J].
AHMAD, F ;
ROBERTS, DJ .
JOURNAL OF CHROMATOGRAPHY A, 1995, 693 (01) :167-175
[2]   Confirmation of conjugation processes during TNT metabolism by axenic plant roots [J].
Bhadra, R ;
Wayment, DG ;
Hughes, JB ;
Shanks, JV .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1999, 33 (03) :446-452
[3]  
BORCH T, 2002, EOS T AGU, V83
[4]   Quantitative structure-activity relationships in enzymatic single-electron reduction of nitroaromatic explosives: implications for their cytotoxicity [J].
Cenas, N ;
Nemeikaite-Ceniene, A ;
Sergediene, E ;
Nivinskas, H ;
Anusevicius, Z ;
Sarlauskas, J .
BIOCHIMICA ET BIOPHYSICA ACTA-GENERAL SUBJECTS, 2001, 1528 (01) :31-38
[5]   Biological treatment of TNT-contaminated soil. 1. Anaerobic cometabolic reduction and interaction of TNT and metabolites with soil components [J].
Daun, G ;
Lenke, H ;
Reuss, M ;
Knackmuss, HJ .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1998, 32 (13) :1956-1963
[6]   Simultaneous variation of temperature and gradient steepness for reversed-phase high-performance liquid chromatography method development - I. Application to 14 different samples using computer simulation [J].
Dolan, JW ;
Snyder, LR ;
Djordjevic, NM ;
Hill, DW ;
Saunders, DL ;
Van Heukelem, L ;
Waeghe, TJ .
JOURNAL OF CHROMATOGRAPHY A, 1998, 803 (1-2) :1-31
[7]   Temperature selectivity in reversed-phase high performance liquid chromatography [J].
Dolan, JW .
JOURNAL OF CHROMATOGRAPHY A, 2002, 965 (1-2) :195-205
[8]  
*EPA, 1994, 8330 EPA, P1
[9]   Determination of nitroaromatic explosives and their degradation products in unsaturated-zone water samples by high-performance liquid chromatography with photodiode-array, mass spectrometric, and tandem mass spectrometric detection [J].
Gates, PM ;
Furlong, ET ;
Dorsey, TF ;
Burkhardt, MR .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1996, 15 (08) :319-325
[10]   Detection of the cyclic nitramine explosives hexahydro-1,3,5-trinitro-1,3,5-triazine (RDX) and octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazine (HMX) and their degradation products in soil environments [J].
Groom, CA ;
Beaudet, S ;
Halasz, A ;
Paquet, L ;
Hawari, J .
JOURNAL OF CHROMATOGRAPHY A, 2001, 909 (01) :53-60