RINGLET motion correction for 3D MRI acquired with the elliptical centric view order

被引:7
作者
Bernstein, MA [1 ]
Shu, YH [1 ]
Elliott, AM [1 ]
机构
[1] Mayo Clin & Mayo Fdn, Rochester, MN 55905 USA
关键词
elliptical centric; 3D; motion correction; RINGLET; marker;
D O I
10.1002/mrm.10584
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
A new rigid-body motion correction algorithm is described that is compatible with 3D image sets acquired with the elliptical centric (EC) view order. With this view order, an annular ring of k-space data is acquired in the ky-kz plane during any short time interval. Images for tracking motion can be reconstructed in the yz-plane from any ring of the acquisition data. In these tracking images, a point source (such as an external marker) shows a characteristic bull's-eye pattern that permits motion monitoring and correction. The true position of the point object is located at the center of the bull's-eye pattern. Cross correlation can be performed to automatically track the positions of markers reconstructed from adjacent rings of k-space. To increase the marker signal, the markers are encased in inductively coupled RF coils. Rigid-body motion in the yz-plane is calculated directly with the Euclidean group for rotation and translation, and corrected by rotating and applying phase shifts to any corrupted rings of data. In the current work we present a theoretical analysis of this method, as well as results of volunteer and controlled phantom experiments that demonstrate its initial feasibility. Although the EC view order has mainly been used for MR angiography (MRA), it can also be used for most 3D acquisitions. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:802 / 812
页数:11
相关论文
共 40 条
[1]
Atkinson D, 1999, MAGNET RESON MED, V41, P163, DOI 10.1002/(SICI)1522-2594(199901)41:1<163::AID-MRM23>3.0.CO
[2]
2-9
[3]
Dynamic scan-plane tracking using MR position monitoring [J].
Derbyshire, JA ;
Wright, GA ;
Henkelman, RM ;
Hinks, RS .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (04) :924-932
[4]
ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES [J].
EHMAN, RL ;
FELMLEE, JP .
RADIOLOGY, 1989, 173 (01) :255-263
[5]
Fain SB, 1999, MAGNET RESON MED, V42, P1106, DOI 10.1002/(SICI)1522-2594(199912)42:6<1106::AID-MRM15>3.0.CO
[6]
2-Q
[7]
A method for fast 3D tracking using tuned fiducial markers and a limited projection reconstruction FISP (LPR-FISP) sequence [J].
Flask, C ;
Elgort, D ;
Wong, E ;
Shankaranarayanan, A ;
Lewin, J ;
Wendt, M ;
Duerk, JL .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (05) :617-627
[8]
ORBITAL NAVIGATOR ECHOES FOR MOTION MEASUREMENTS IN MAGNETIC-RESONANCE-IMAGING [J].
FU, ZW ;
WANG, Y ;
GRIMM, RC ;
ROSSMAN, PJ ;
FELMLEE, JP ;
RIEDERER, SJ ;
EHMAN, RL .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (05) :746-753
[9]
Correction of motion artifacts in linogram and projection reconstruction MRI using geometry and consistency constraints [J].
Gai, N ;
Axel, L .
MEDICAL PHYSICS, 1996, 23 (02) :251-262
[10]
CONSISTENT PROJECTION RECONSTRUCTION (CPR) TECHNIQUES FOR MRI [J].
GLOVER, GH ;
NOLL, DC .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (03) :345-351