International comparison of radiation-temperature measurements with filtered detectors over the temperature range 1380 K to 3100 K

被引:12
作者
Harrison, NJ [1 ]
Fox, NP
Sperfeld, P
Metzdorf, J
Khlevnoy, BB
Stolyarevskaya, RI
Khromchenko, VB
Mekhontsev, SN
Shapoval, VI
Zelener, MF
Sapritsky, VI
机构
[1] Natl Phys Lab, Ctr Opt & Environm Metrol, Teddington TW11 0LW, Middx, England
[2] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
[3] All Russian Res Inst Optophys Measurements, VNIIOFI, Moscow 119361, Russia
关键词
D O I
10.1088/0026-1394/35/4/12
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A series of black-body radiation-temperature measurements has been made over the temperature range 1380 K to 3100 K using two different designs of pyrolytic-graphite black bodies with calculated emissivities of 0.999. All measurements were performed at the All-Russian Research Institute for Optophysical Measurements (VNIIOFI) during June 1997. A filter photometer from the VNIIOFI, broadband glass-filter detectors from the Physikalisch-Technische Bundesanstalt (PTB) and narrowband interference-filter-based radiometers from the National Physical Laboratory (NPL) were used to perform the temperature measurements in either radiance or irradiance mode using both black bodies. Across the entire temperature range, the NPL and PTB instruments showed consistent results with both black bodies and differing geometrical arrangements. Results from the VNIIOFI photometer were also broadly consistent for a wide temperature range.
引用
收藏
页码:283 / 288
页数:6
相关论文
共 9 条
[1]   HIGHLY STABLE, MONOCHROMATIC AND TUNABLE OPTICAL RADIATION SOURCE AND ITS APPLICATION TO HIGH-ACCURACY SPECTROPHOTOMETRY [J].
ANDERSON, VE ;
FOX, NP ;
NETTLETON, DH .
APPLIED OPTICS, 1992, 31 (04) :536-545
[2]   ABSOLUTE SPECTRAL RADIOMETRIC DETERMINATION OF THE THERMODYNAMIC TEMPERATURES OF THE MELTING FREEZING POINTS OF GOLD, SILVER AND ALUMINUM [J].
FOX, NP ;
MARTIN, JE ;
NETTLETON, DH .
METROLOGIA, 1991, 28 (05) :357-374
[3]   High-accuracy characterisation and applications of filter radiometers [J].
Fox, NP ;
Chunnilall, CJ ;
Harrison, NJ ;
Hartree, WS .
OPTICAL RADIATION MEASUREMENTS III, 1996, 2815 :32-41
[4]   A METHOD OF REALIZING SPECTRAL IRRADIANCE BASED ON AN ABSOLUTE CRYOGENIC RADIOMETER [J].
JOHNSON, BC ;
CROMER, CL ;
SAUNDERS, RD ;
EPPELDAUER, G ;
FOWLER, J ;
SAPRITSKY, VI ;
DEZSI, G .
METROLOGIA, 1993, 30 (04) :309-315
[5]  
METZDORF J, 1996, ADV METROLOGY ITS RO, P297
[6]  
SAPRITSKY VI, IN PRESS UTILIZATION
[7]   Spectral-irradiance scale based on radiometric black-body temperature measurements [J].
Sperfeld, P ;
Raatz, KH ;
Nawo, B ;
Moller, W ;
Metzdorf, J .
METROLOGIA, 1996, 32 (06) :435-439
[8]   Investigation of high-temperature black body BB3200 [J].
Sperfeld, P ;
Metzdorf, J ;
Harrison, NJ ;
Fox, NP ;
Khlevnoy, BB ;
Khromchenko, VB ;
Mekhontsev, SN ;
Shapoval, VI ;
Zelener, MF ;
Sapritsky, VI .
METROLOGIA, 1998, 35 (04) :419-422
[9]  
WHITE N, 1995, METROLOGIA, V32, P431