X-ray characterization of breast phantom materials

被引:65
作者
Byng, JW [1 ]
Mainprize, JG [1 ]
Yaffe, MJ [1 ]
机构
[1] Univ Toronto & Imaging Res, Sunnybrook Hlth Sci Ctr, Dept Med Biophys & Radiol, Toronto, ON M4N 3M5, Canada
关键词
D O I
10.1088/0031-9155/43/5/026
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A pulse-height spectroscopic technique is used to measure the linear attenuation coefficients of commercially available composite phantom materials designed to simulate the attenuation characteristics of breast fat and breast glandular tissue. The manufacturers have specified the composition of these materials with the goal of matching the linear attenuation coefficients of breast tissues, calculated using the mixture rule. Over the energy range 18 to 100 keV, measurements from these materials are in close agreement with manufacturers' predictions and with previously measured linear attenuation coefficients of breast tissue samples.
引用
收藏
页码:1367 / 1377
页数:11
相关论文
共 13 条
[1]   DEVELOPMENT OF AN ANTHROPOMORPHIC BREAST PHANTOM [J].
CALDWELL, CB ;
YAFFE, MJ .
MEDICAL PHYSICS, 1990, 17 (02) :273-280
[2]  
*CIRS, 1993, TECHN PAP TISS EQ PH
[3]  
HAMMERSTEIN GR, 1979, RADIOLOGY, V130, P485
[4]   PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV [J].
HUBBELL, JH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1269-1290
[5]  
ICRP ( International Commission On Radiological Protection, 1975, 23 ICRP, V23
[6]   X-RAY CHARACTERIZATION OF NORMAL AND NEOPLASTIC BREAST TISSUES [J].
JOHNS, PC ;
YAFFE, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (06) :675-695
[7]   CORRECTION OF PULSE-HEIGHT SPECTRA FOR PEAK PILEUP EFFECTS USING PERIODIC AND RANDOM PULSE GENERATORS [J].
JOHNS, PC ;
YAFFE, MJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 255 (03) :559-581
[8]   GENERALIZED IMAGE COMBINATIONS IN DUAL KVP DIGITAL RADIOGRAPHY [J].
LEHMANN, LA ;
ALVAREZ, RE ;
MACOVSKI, A ;
BRODY, WR ;
PELC, NJ ;
RIEDERER, SJ ;
HALL, AL .
MEDICAL PHYSICS, 1981, 8 (05) :659-667
[9]  
Plechaty E F, 1975, TABLES GRAPHS PHOTON, V6
[10]   FORMULATION OF TISSUE SUBSTITUTE MATERIALS USING BASIC INTERACTION DATA [J].
WHITE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (05) :889-899