Influence. of electrode misplacement on the electrocardiographic signs of inferior myocardial ischemia

被引:29
作者
Rudiger, A
Schöb, L
Follath, F
机构
[1] Univ Zurich Hosp, Dept Internal Med, CH-8091 Zurich, Switzerland
[2] Ita Wegmann Clin, Arlesheim, Switzerland
关键词
ECG; artifact; electrode; misplacement;
D O I
10.1016/j.ajem.2003.08.007
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Electrocardographic (ECG) artifacts resulting from misplacements of electrodes are frequent, difficult to detect, and can become of clinical importance. We investigated 2 healthy volunteers and 3 patients with ECG signs of inferior myocardial scars. We exchanged the peripheral electrodes in a defined manner and investigated the resulting ECG for morphology and possible diagnostic errors. In the volunteers, ECG signs of inferior ischemia could be produced. In the patients with ischemic heart disease, normal ECG without signs of ischemia resulted by placing the electrode of the left leg to the left arm. The automatic ECG analyzer was not helpful in detecting artifacts by misplaced electrodes. A very low amplitude of the QRS complex in lead I, II, or III was pathognomonic for electrode misplacement in half of the cases. ECG artifacts must also be suspected when abnormal QRS- or P-axis occur or when QRS morphology does not match with the clinical presentation of the patient. (Am J Emerg Med 2003;21:574-577. (C) 2003 Elsevier Inc. All rights reserved.).
引用
收藏
页码:574 / 577
页数:4
相关论文
共 11 条
  • [1] β-blockers and reduction of cardiac events in noncardiac surgery -: Clinical applications
    Auerbach, AD
    Goldman, L
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2002, 287 (11): : 1445 - 1447
  • [2] BRAUNWALD E, 1997, TXB CARDIOVASCULAR M, P141
  • [3] ABNORMAL ECG IN CLINICALLY NORMAL INDIVIDUALS
    FISCH, C
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1983, 250 (10): : 1321 - 1323
  • [4] ECG DIAGNOSTIC ERRORS DUE TO IMPROPER CONNECTION OF THE RIGHT ARM AND LEG CABLES
    GUIJARROMORALES, A
    GILEXTREMERA, B
    MALDONADOMARTIN, A
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 1991, 30 (02) : 233 - 235
  • [5] RECOGNITION OF ELECTROCARDIOGRAPHIC ELECTRODE MISPLACEMENTS INVOLVING THE GROUND (RIGHT LEG) ELECTRODE
    HAISTY, WK
    PAHLM, O
    EDENBRANDT, L
    NEWMAN, K
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (16) : 1490 - 1495
  • [6] Detection of frequently overlooked electrocardiographic lead reversals using artificial neural networks
    Heden, B
    Ohlsson, M
    Holst, H
    Mjoman, M
    Rittner, R
    Pahlm, O
    Peterson, C
    Edenbrandt, L
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (05) : 600 - &
  • [7] ARTIFACT AND SAFETY IN ECG MONITORING
    JONES, PL
    [J]. PROCEEDINGS OF THE ROYAL SOCIETY OF MEDICINE-LONDON, 1973, 66 (12): : 1211 - 1215
  • [8] RECOGNITION OF ELECTROCARDIOGRAPHIC LEAD MISPLACEMENTS
    PEBERDY, MA
    ORNATO, JP
    [J]. AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1993, 11 (04) : 403 - 405
  • [9] Rudiger A, 2001, SWISS MED FORUM, V28, P741
  • [10] TUNA N, 2000, CARDIOVASCULAR MED, P29