Monte Carlo study of TLD measurements in air cavities

被引:12
作者
Haraldsson, P
Knöö, T
Nyström, H [1 ]
Engström, P
机构
[1] Univ Copenhagen Hosp, Finsen Ctr, Dept Radiat Phys, DK-2100 Copenhagen, Denmark
[2] Univ Lund Hosp, SE-22185 Lund, Sweden
关键词
D O I
10.1088/0031-9155/48/18/401
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Thermoluminescent dosimeters (TLDs) are used for verification of the delivered dose during IMRT treatment of head and neck carcinomas. The TLDs are put into a plastic tube, which is placed in the nasal cavities through the treated volume. In this study, the dose distribution to a phantom having a cylindrical air cavity containing a tube was calculated by Monte Carlo methods and the results were compared with data from a treatment planning system (TPS) to evaluate the accuracy of the TLD measurements. The phantom was defined in the DOSXYZnrc Monte Carlo code and calculations were performed with 6 MV fields, with the TLD tube placed at different positions within the cylindrical air cavity. A similar phantom was defined in the pencil beam based TPS. Differences between the Monte Carlo and the TPS calculations of the absorbed dose to the TLD tube were found to be small for an open symmetrical field. For a half-beam field through the air cavity, there was a larger discrepancy. Furthermore, dose profiles through the cylindrical air cavity show, as expected, that the treatment planning system overestimates the absorbed dose in the air cavity. This study shows that when using an open symmetrical field, Monte Carlo calculations of absorbed doses to a TLD tube in a cylindrical air cavity give results comparable to a pencil beam based treatment planning system.
引用
收藏
页码:N253 / N259
页数:7
相关论文
共 8 条
[1]   Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version [J].
Kawrakow, I .
MEDICAL PHYSICS, 2000, 27 (03) :485-498
[2]  
KHAN F, 1994, PHYSICS RADIATION TH
[3]   LIMITATIONS OF A PENCIL BEAM APPROACH TO PHOTON DOSE CALCULATIONS IN LUNG-TISSUE [J].
KNOOS, T ;
AHNESJO, A ;
NILSSON, P ;
WEBER, L .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (09) :1411-1420
[4]   A systematic evaluation of air cavity dose perturbation in megavoltage x-ray beams [J].
Li, XA ;
Yu, C ;
Holmes, T .
MEDICAL PHYSICS, 2000, 27 (05) :1011-1017
[5]  
Walters B. R. B., 2002, DOSXYZNRC USERS MANU
[6]  
WALTERS BRB, 2001, BEAMNRC USERS MANUAL
[7]   Monte Carlo evaluation of tissue inhomogeneity effects in the treatment of the head and neck [J].
Wang, L ;
Yorke, E ;
Chui, CS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (05) :1339-1349
[8]   A virtual linear accelerator for verification of treatment planning systems [J].
Wieslander, E ;
Knoos, T .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (10) :2887-2896