Electrochemical detection of catecholamines at sub-5 fg levels by redox cycling

被引:12
作者
Cullison, JK [1 ]
Waraska, J [1 ]
Buttaro, DJ [1 ]
Acworth, IN [1 ]
Bowers, ML [1 ]
机构
[1] ESA, Chelmsford, MA 01842 USA
关键词
electrochemical detection; catecholamines; neurotransmitters; redox cycling; low injection volumes;
D O I
10.1016/S0731-7085(98)00136-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Two simple modifications to a commercially available thin layer electrochemical detector cell permitted the attainment of ultra-low detection levels of two neurotransmitter catecholamines. An ESA model 5041 analytical cell was modified with a glassy carbon embedded ceramic composite electrode to allow the use of a thin 12 mu m gasket. Also a capillary HPLC column was connected directly to the detector cell using a 30 mu m i.d. fused silica capillary. These modifications permitted the extensive redox cycling of the electrochemically reversible catecholamines. The ensuing amplified analytical signal allowed the detector cell to achieve efficiencies of 1300%. This resulted in a mass limit of detection of 4 fg and a concentration limit of detection of 116 pM for dopamine with an S/N of 3. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:253 / 259
页数:7
相关论文
共 11 条
[1]  
ACWORTH IN, 1997, PROGR HPLC, V6, P3
[2]  
ANDREWS RW, 1982, AM LAB, V14, P140
[3]  
BARD AJ, 1980, ELECTROCHEMICAL METH, P7
[4]   A microfabricated flow-through cell with parallel-opposed electrodes for recycling amperometric detection [J].
Brooks, SA ;
Kennedy, RT .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1997, 436 (1-2) :27-34
[5]  
DeDeurwaerdere P, 1997, J NEUROCHEM, V68, P195
[6]  
DUDA CT, 1993, TECHNIQUES BEHAV NEU, V11, P50
[7]   LIQUID-CHROMATOGRAPHY DETECTOR BASED ON SINGLE AND TWIN ELECTRODE THIN-LAYER ELECTROCHEMISTRY - APPLICATION TO DETERMINATION OF CATECHOLAMINES IN BLOOD-PLASMA [J].
FENN, RJ ;
SIGGIA, S ;
CURRAN, DJ .
ANALYTICAL CHEMISTRY, 1978, 50 (08) :1067-1073
[8]   ROLE OF COLUMN PARAMETERS AND INJECTION VOLUME ON DETECTION LIMITS IN LIQUID-CHROMATOGRAPHY [J].
KARGER, BL ;
MARTIN, M ;
GUIOCHON, G .
ANALYTICAL CHEMISTRY, 1974, 46 (12) :1640-1647
[9]  
Kissinger P. T., 1991, MICRODIALYSIS NEUROS, P103
[10]   NOISE AND SIGNAL-TO-NOISE RATIO IN ELECTROCHEMICAL DETECTORS [J].
MORGAN, DM ;
WEBER, SG .
ANALYTICAL CHEMISTRY, 1984, 56 (13) :2560-2567