Response of an iridium oxide pH-sensor in nonaqueous solutions. Comparison with other pH-sensors

被引:52
作者
Izutsu, K
Yamamoto, H
机构
[1] Faculty of Science, Shinshu University, Asahi
关键词
iridium oxide pH-sensor; pH-ISFET; glass electrode; nonaqueous solvent; pH of nonaqueous solutions; potentiometry;
D O I
10.2116/analsci.12.905
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The response of an iridium oxide (IrO2) pH-sensor (nonaqueous version of METOX, TOA Electronics Ltd.) was studied in various dipolar aprotic solvents. From the potential measurements in picric acid-picrate solutions, it was confirmed to respond in a Nernstian way to the pH in all the solvents studied. The response speed of the IrO2 pH-sensor was compared with that of the glass electrode as well as those of the Si3N4-and Ta2O4-type pH-ISFETs, carrying out acid-base titrations at various speeds of titrant addition. The response speed of the IrO2 pH-sensor was much higher than that of the glass electrode, though it was somewhat lower than those of the pH-ISFETs. The IrO2 pH-sensor responded in a wide pH range, gave reproducible results, needed no pretreatment, and was durable over a year. It is promising for pH measurements in various nonaqueous solutions free from redox systems.
引用
收藏
页码:905 / 909
页数:5
相关论文
共 14 条
[1]  
[Anonymous], 1990, IUPAC CHEM DATA SERI
[2]  
BATES RG, 1973, DETERMINATION PH THE, P372
[3]   IRIDIUM IRIDIUM OXIDE ELECTRODE FOR POTENTIOMETRIC DETERMINATION OF PROTON ACTIVITY IN HYDROORGANIC SOLUTIONS AT SUB-ZERO TEMPERATURES [J].
BORDI, S ;
CARLA, M ;
PAPESCHI, G ;
PINZAUTI, S .
ANALYTICAL CHEMISTRY, 1984, 56 (02) :317-319
[4]  
Coetzee J. F, 1982, RECOMMENDED METHODS
[5]  
Coetzee J. M., 2005, COMMUNICATION
[6]   ACID-BASE EQUILIBRIA OF SOME ACIDS IN PROPYLENE CARBONATE [J].
IZUTSU, K ;
KOLTHOFF, IM ;
FUJINAGA, T ;
HATTORI, M ;
CHANTOONI, MK .
ANALYTICAL CHEMISTRY, 1977, 49 (03) :503-508
[7]   Acid-base equilibria in gamma-butyrolactone studied by use of pH-ISFETs [J].
Izutsu, K ;
Ohmaki, M .
TALANTA, 1996, 43 (04) :643-648
[8]   USE OF PH-SENSITIVE ISFETS AS SENSORS FOR PH IN NONAQUEOUS SOLUTIONS AND FOR PROTON SOLVATION [J].
IZUTSU, K ;
NAKAMURA, T ;
HIRAOKA, S .
CHEMISTRY LETTERS, 1993, (11) :1843-1846
[9]   SOME RECENT STUDIES ON THE USE OF ELECTROCHEMICAL SENSORS IN NONAQUEOUS SOLUTION CHEMISTRY [J].
IZUTSU, K ;
NAKAMURA, T ;
ARAI, T ;
OHMAKI, M .
ELECTROANALYSIS, 1995, 7 (09) :884-888
[10]   CHARACTERIZATION OF OXIDE-FILMS ELECTROCHEMICALLY DEPOSITED FROM SOLUTIONS OF PALLADIUM-CHLORIDE AND SODIUM HEXACHLOROIRIDATE [J].
JAWORSKI, RK ;
COX, JA ;
STROHMEIER, BR .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1992, 325 (1-2) :111-123