Prospective comparison of temperature guided microwave and radiofrequency catheter ablation in the swine heart

被引:13
作者
Jumrussirikul, P
Chen, JT
Jenkins, M
Hui, R
Taylor, K
Wang, PJ
Hutchins, GM
Calkins, H
机构
[1] Johns Hopkins Med Inst, Dept Med, Baltimore, MD 21205 USA
[2] Johns Hopkins Med Inst, Dept Pathol, Baltimore, MD 21205 USA
[3] Tufts Univ, Sch Med, Dept Med, Medford, MA 02155 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1998年 / 21卷 / 07期
关键词
catheter ablation; microwave energy; radiofrequency energy;
D O I
10.1111/j.1540-8159.1998.tb00206.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Microwave energy has been proposed as an alternative to radiofrequency energy for use during catheter ablation procedures. The purpose of this study was to prospectively compare, in an animal model, the lesion size associated with temperature guided catheter ablation using either microwave or radiofrequency energy. Eleven swine underwent catheter ablation with either radiofrequency (N = 4) or microwave energy (N = 7). In each animal catheter ablation was performed at 7-15 sites. At each site energy was delivered for 60 seconds using closed loop feedback temperature control to achieve a target temperature of 70 degrees C. Cardiac catheterization was performed before and after ablation. Animals were sacrificed approximately one month following the ablation procedure. Analysis of lesion size demonstrated that overall lesions created using radiofrequency energy were larger than those created using microwave energy. In the ventricle, lesions created using microwave energy were longer, but had a similar width and depth as those created using radiofrequency energy. An important relation was observed between the depth of lesions created using microwave energy and catheter stability, as evidenced by the temperature profile. Overall, lesions created using microwave energy are smaller than those created using radiofrequency energy. Catheter stability has an important impact on lesion size.
引用
收藏
页码:1364 / 1374
页数:11
相关论文
共 38 条
[1]   ASSESSMENT OF EFFECTS OF A RADIOFREQUENCY ENERGY-FIELD AND THERMISTOR LOCATION IN AN ELECTRODE CATHETER ON THE ACCURACY OF TEMPERATURE-MEASUREMENT [J].
BLOUIN, LT ;
MARCUS, FI ;
LAMPE, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (05) :807-813
[2]   CATHETER ABLATION OF ATRIAL-FLUTTER USING RADIOFREQUENCY ENERGY [J].
CALKINS, H ;
LEON, AR ;
DEAM, AG ;
KALBFLEISCH, SJ ;
LANGBERG, JJ ;
MORADY, F .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 73 (05) :353-356
[3]   RADIOFREQUENCY CATHETER ABLATION OF ACCESSORY ATRIOVENTRICULAR CONNECTIONS IN 250 PATIENTS - ABBREVIATED THERAPEUTIC APPROACH TO WOLFF-PARKINSON-WHITE SYNDROME [J].
CALKINS, H ;
LANGBERG, J ;
SOUSA, J ;
ELATASSI, R ;
LEON, A ;
KOU, W ;
KALBFLEISCH, S ;
MORADY, F .
CIRCULATION, 1992, 85 (04) :1337-1346
[4]   RADIOFREQUENCY ABLATION OF THE INFERIOR VENA-CAVA - TRICUSPID-VALVE ISTHMUS IN COMMON ATRIAL-FLUTTER [J].
COSIO, FG ;
LOPEZGIL, M ;
GOICOLEA, A ;
ARRIBAS, F ;
BARROSO, JL .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (08) :705-709
[5]   RADIOFREQUENCY CATHETER ABLATION FOR THE TREATMENT OF HUMAN TYPE-1 ATRIAL-FLUTTER - IDENTIFICATION OF A CRITICAL ZONE IN THE REENTRANT CIRCUIT BY ENDOCARDIAL MAPPING TECHNIQUES [J].
FELD, GK ;
FLECK, RP ;
CHEN, PS ;
BOYCE, K ;
BAHNSON, TD ;
STEIN, JB ;
CALISI, CM ;
IBARRA, M .
CIRCULATION, 1992, 86 (04) :1233-1240
[6]  
GADHOKE A, 1992, CIRCULATION, V86, P912
[7]   ELECTRODE RADIUS PREDICTS LESION RADIUS DURING RADIOFREQUENCY ENERGY HEATING - VALIDATION OF A PROPOSED THERMODYNAMIC MODEL [J].
HAINES, DE ;
WATSON, DD ;
VEROW, AF .
CIRCULATION RESEARCH, 1990, 67 (01) :124-129
[8]   OBSERVATIONS ON ELECTRODE-TISSUE INTERFACE TEMPERATURE AND EFFECT ON ELECTRICAL-IMPEDANCE DURING RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
HAINES, DE ;
VEROW, AF .
CIRCULATION, 1990, 82 (03) :1034-1038
[9]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[10]   RADIOFREQUENCY CATHETER ABLATION IN UNUSUAL MECHANISMS OF ATRIAL-FIBRILLATION - REPORT OF 3 CASES [J].
HAISSAGUERRE, M ;
MARCUS, FI ;
FISCHER, B ;
CLEMENTY, J .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1994, 5 (09) :743-751