Concept, design and fabrication of smart orthopedic implants

被引:73
作者
Burny, F
Donkerwolcke, M
Moulart, F
Bourgois, R
Puers, R
Van Schuylenbergh, K
Barbosa, M
Paiva, O
Rodes, F
Bégueret, JB
Lawes, P
机构
[1] Free Univ Brussels, Hop Erasme, Clin Univ Bruxelles, Serv Orthopedie Traumatol, B-1070 Brussels, Belgium
[2] Royal Mil Acad, Brussels, Belgium
[3] Katholieke Univ Leuven, ESAT MICAS, Louvain, Belgium
[4] Inst Biomed Engn, INEB, Dept Met, Porto, Portugal
[5] Univ Porto, Fac Engn, P-4100 Porto, Portugal
[6] Inst Politecn Porto, Oporto, Portugal
[7] Univ Bordeaux 1, IXL, Bordeaux, France
[8] Howmed Int Inc, Raheen, Ireland
关键词
failure of implants; fracture healing; orthopedic implants; rehabilitation; sensors; smart implants; strain measurements; telemetry;
D O I
10.1016/S1350-4533(00)00062-X
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Extensive clinical experience has been built up using orthopaedic implants instrumented with strain gauges connected to a Wheatstone bridge by means of percutaneous leads. This research showed the medical relevance of the monitoring of the deformation of implants as a powerful tool to evaluate nursing and rehabilitation exercises, for tracing dangerous overloads and anticipating implant failure and also to observe the healing process. The IMPACT 3500 project focuses on the instrumentation of femoral implants with on board sensors: regular Benoist-Girard implants have been modified. to contain a 'sensing cell', and thoroughly tested in vitro and in vivo. The implant deformations are measured with resistive strain gauges, and the signal is transferred to a personal computer for processing and display, via a hard wired connection, or via a telemetry system. Two fully implantable wireless designs, called Linkstrain and Sealstrain, are powered from the outside by magnetic induction. As Sealstrain contains the whole telemetric system in its cavity, the highest miniaturization was required; this seriously deteriorates the efficiency of the inductive power link. (C) 2001 IPEM. Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:469 / 479
页数:11
相关论文
共 19 条
[1]  
BARBOSA MA, 1992, BIOMATERIALS HARD TI, P257
[2]  
Burny F, 1971, Acta Orthop Belg, V37, P266
[3]  
Burny F, 1976, Acta Orthop Belg, V42, P31
[4]  
BURNY F, 1991, CONCISE ENCY BIOL BI, P391
[5]  
BURNY F, 1993, MONITORING ORTHOPEDI, P138
[6]   ANALYSIS OF RESONANT COUPLED COILS IN THE DESIGN OF RADIO-FREQUENCY TRANS-CUTANEOUS LINKS [J].
DONALDSON, ND ;
PERKINS, TA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1983, 21 (05) :612-627
[7]  
DONKERWOLCKE M, 1993, MONITORING ORTHOPEDI, P39
[8]   MUTUAL INDUCTANCE OF AIR-CORED COILS - EFFECT ON DESIGN OF RADIO-FREQUENCY COUPLED IMPLANTS [J].
FLACK, FC ;
JAMES, ED ;
SCHLAPP, DM .
MEDICAL & BIOLOGICAL ENGINEERING, 1971, 9 (02) :79-&
[9]   MACHINING A MACHINABLE GLASS-CERAMIC [J].
GROSSMAN, DG .
VACUUM, 1978, 28 (02) :55-61
[10]   ENERGIZING IMPLANTABLE TRANSMITTERS BY MEANS OF COUPLED INDUCTANCE COILS [J].
KADEFORS, R ;
KAISER, E ;
PETERSEN, I .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1969, BM16 (03) :177-&