An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetry

被引:92
作者
Oldham, M
Baustert, I
Lord, C
Smith, TAD
McJury, M
Warrington, AP
Leach, MO
Webb, S
机构
[1] Inst Canc Res, Joint Dept Phys, Sutton SM2 5PT, Surrey, England
[2] Royal Marsden NHS Trust, Sutton SM2 5PT, Surrey, England
[3] Inst Canc Res, CRC, Clin MR Res Grp, Sutton SM2 5PT, Surrey, England
[4] Royal Marsden NHS Trust, Sutton SM2 5PT, Surrey, England
[5] Inst Canc Res, Dept Nucl Med, Sutton SM2 5PT, Surrey, England
关键词
D O I
10.1088/0031-9155/43/5/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
BANG-gel dosimetry offers the potential for measuring the dose delivered by a radiotherapy treatment technique, in three dimensions, with high spatial resolution and good accuracy. The ability to measure comprehensively a 3D dose distribution is a major advantage of the gel dosimeter over conventional planar and point-based dosimeter devices, particularly when applied to the verification of complex dose distributions characteristic of intensity-modulated radiotherapy (IMRT). In this paper an in-house manufactured BANG-gel dosimeter was applied to study the dose distributions of two irradiation experiments for which the distributions were known: (i) a dosimetrically simple parallel-opposed irradiation, and (ii) a more complex nine-field 'static tomotherapy' intensity-modulated irradiation delivered with the Nomos MIMiC. The uniform distribution in (i) allowed a study of the magnetic resonance (MR) imaging parameters to achieve an optimal trade-off between noise and image resolution (optimum image resolution for the Siemens 1.5T Vision system was determined to be approximate to 0.8 mm(2) with a slice thickness of 2 mm). The spatial uniformity of gel sensitivity to radiation was found to depend strongly on the presence of oxygen, which must be eliminated for the gel dosimeter to be of use. The gel dosimeter was found to agree well with predicted dose distributions and accurately measured the steep penumbral fall-off of dose, even after many days, proving its potential for the verification of IMRT distributions. In the nine-field IMRT delivery (ii) the predicted dose was computed by both an in-house 'component-delivery' dose algorithm and the Peacock planning-system dose algorithm. Good agreement was found between the two algorithms despite the latter's omission of the change in penumbral characteristics with aperture-size during delivery, lack of inhomogeneity correction and approximate modelling of leaf leakage. These effects were found to be small for the problem studied. The predicted distribution agreed well with the gel-measured distribution at medium and high doses (50-90% isodose lines) although differences of up to 10% were observed at lower doses (30% isodose line). The gel dosimeter was found to have the potential to verify IMRT distributions but required considerable care to achieve accurate results. Attention was required to achieve uniformity of gel sensitivity (to prevent oxygen contamination), and in the calibration process.
引用
收藏
页码:1113 / 1132
页数:20
相关论文
共 22 条
[1]   DIFFUSION IN FE(II/III) RADIATION-DOSIMETRY GELS MEASURED BY MAGNETIC-RESONANCE-IMAGING [J].
BALCOM, BJ ;
LEES, TJ ;
SHARP, AR ;
KULKARNI, NS ;
WAGNER, GS .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (10) :1665-1676
[2]   CLINICAL REALIZATION OF 3D CONFORMAL INTENSITY-MODULATED RADIOTHERAPY [J].
BUTLER, EB ;
WOO, SY ;
GRANT, W ;
NIZIN, PS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 32 (05) :1547-1548
[3]   The field-matching problem as it applies to the peacock three dimensional conformal system for intensity modulation [J].
Carol, M ;
Grant, WH ;
Bleier, AR ;
Kania, AA ;
Targovnik, HS ;
Butler, EB ;
Woo, SW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (01) :183-187
[4]   PEACOCK(TM) - A SYSTEM FOR PLANNING AND ROTATIONAL DELIVERY OF INTENSITY-MODULATED FIELDS [J].
CAROL, MP .
INTERNATIONAL JOURNAL OF IMAGING SYSTEMS AND TECHNOLOGY, 1995, 6 (01) :56-61
[5]  
CAROL MP, 1994, USE COMPUTERS RAD TH, P108
[6]  
CAROL MP, 1994, USE COMPUTERS RAD TH, P172
[7]  
CAROL MP, 1993, 3 DIMENSIONAL TREATM, P173
[8]   Dosimetry of high energy photon and electron beams with CEA films [J].
Cheng, CW ;
Das, IJ .
MEDICAL PHYSICS, 1996, 23 (07) :1225-1232
[9]  
FARAJOLLAHI AR, 1997, P WORLD C MED PHYSIC, V2
[10]  
Fricke H, 1966, RAD DOSIMETRY, V2