Back-projection algorithm with misalignment corrections for 2D3C stereoscopic PIV

被引:65
作者
Coudert, SJM [1 ]
Schon, JP [1 ]
机构
[1] Univ St Etienne, Lab Traitement Signal & Instrumentat, UMR 5516, CNRS, F-42023 St Etienne Du Rouvray 2, France
关键词
stereoscopic PIV; mapping; warping; misalignment; corrections; Scheimpflug condition; cross-correlation; measurement accuracy;
D O I
10.1088/0957-0233/12/9/301
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
in stereoscopic PIV mapping and warping algorithms are used to back-project the image information on the object plane. The mapping algorithm back-projects each pixel, whereas the warping algorithm back-projects the 2D displacement vectors. In this paper, both algorithms and the misalignment of the laser sheet regarding the object plane are considered. Then, a new algorithm using both mapping and warping algorithms is described for a 2D3C angular stereoscopic PIV set-up with standard lenses and the so-called Scheimpflug condition. The accuracy of the new algorithm, that takes into account laser sheet misalignment, is compared to the previous algorithm on the same sets of images. These sets are recorded from a paper pattern mounted on a 3D translation and rotation stage and from a turbulent pipe flow. The tests show that the accuracy benefit is one order of magnitude on these sets. The accuracy gain depends on the laser sheet misalignment magnitude, which can be measured using the mapping algorithm.
引用
收藏
页码:1371 / 1381
页数:11
相关论文
共 12 条
[1]  
COUDERT S, 2000, EUROMECH, V411
[2]  
GAUTHIER V, 1988, VKI LS 1988 06 PARTI
[3]   Three-dimensional particle image velocimetry: experimental error analysis of a digital angular stereoscopic system [J].
Lawson, NJ ;
Wu, J .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1997, 8 (12) :1455-1464
[4]   Stereoscopic particle image velocimetry [J].
Prasad, AK .
EXPERIMENTS IN FLUIDS, 2000, 29 (02) :103-116
[5]   SCHEIMPFLUG STEREOCAMERA FOR PARTICLE IMAGE VELOCIMETRY IN LIQUID FLOWS [J].
PRASAD, AK ;
JENSEN, K .
APPLIED OPTICS, 1995, 34 (30) :7092-7099
[6]  
PRASAD AK, 1993, EXP FLUIDS, V15, P49, DOI 10.1007/BF00195595
[7]  
RIOU L, 1999, PHYSICS SIGNAL IMAGE
[8]   Distortion compensation for generalized stereoscopic particle image velocimetry [J].
Soloff, SM ;
Adrian, RJ ;
Liu, ZC .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1997, 8 (12) :1441-1454
[9]  
VANOORD J, 1997, 161 MEAH
[10]   The effect of a discrete window offset on the accuracy of cross-correlation analysis of digital PIV recordings [J].
Westerweel, J ;
Dabiri, D ;
Gharib, M .