Variations in PET/CT Methodology for Oncologic Imaging at US Academic Medical Centers: An Imaging Response Assessment Team Survey

被引:74
作者
Graham, Michael M. [1 ]
Badawi, Ramsey D. [2 ]
Wahl, Richard L. [3 ]
机构
[1] Univ Iowa, Dept Radiol JPP 3863, Iowa City, IA 52242 USA
[2] Univ Calif Davis, Med Ctr, Sacramento, CA 95817 USA
[3] Johns Hopkins Univ, Sch Med, Baltimore, MD USA
关键词
oncology; PET/CT; quality assurance; IRAT; PET-CT; standardization; variation; POSITRON-EMISSION-TOMOGRAPHY; LUNG-CANCER; RANDOMIZED-TRIAL;
D O I
10.2967/jnumed.109.074104
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In 2005, 8 Imaging Response Assessment Teams (IRATs) were funded by the National Cancer Institute (NCI) as supplemental grants to existing NCI Cancer Centers. After discussion among the IRATs regarding the need for increased standardization of clinical and research PET/CT methodology, it became apparent that data acquisition and processing approaches differ considerably among centers. To determine the variability in detail, a survey of IRAT sites and IRAT affiliates was performed. Methods: A 34-question instrument evaluating patient preparation, scanner type, performance approach, display, and analysis was developed. Fifteen institutions, including the 8 original IRATs and 7 institutions that had developed affiliate IRATs, were surveyed. Results: The major areas of variation were (18)F-FDG dose (259-740 MBq [7-20 mCi]) uptake time (45-90 min), sedation (never to frequently), handling of diabetic patients, imaging time (2-7 min/bed position), performance of diagnostic CT scans as a part of PET/CT, type of acquisition (2-dimensional vs. 3-dimensional), CT technique, duration of fasting (4 or 6 h), and (varying widely) acquisition, processing, display, and PACS software-with 4 sites stating that poor-quality images appear on PACS. Conclusion: There is considerable variability in the way PET/CT scans are performed at academic institutions that are part of the IRAT network. This variability likely makes it difficult to quantitatively compare studies performed at different centers. These data suggest that additional standardization in methodology will be required so that PET/CT studies, especially those performed quantitatively, are more comparable across sites.
引用
收藏
页码:311 / 317
页数:7
相关论文
共 9 条
[1]  
*ACRIN PET COR LAB, AM COLL RAD IM NETW
[2]   Weight-Based, Low-Dose Pediatric Whole-Body PET/CT Protocols [J].
Alessio, Adam M. ;
Kinahan, Paul E. ;
Manchanda, Vivek ;
Ghioni, Victor ;
Aldape, Lisa ;
Parisi, Marguerite T. .
JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (10) :1570-1578
[3]   FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0 [J].
Boellaard, Ronald ;
O'Doherty, Mike J. ;
Weber, Wolfgang A. ;
Mottaghy, Felix M. ;
Lonsdale, Markus N. ;
Stroobants, Sigrid G. ;
Oyen, Wim J. G. ;
Kotzerke, Joerg ;
Hoekstra, Otto S. ;
Pruim, Jan ;
Marsden, Paul K. ;
Tatsch, Klaus ;
Hoekstra, Corneline J. ;
Visser, Eric P. ;
Arends, Bertjan ;
Verzijlbergen, Fred J. ;
Zijlstra, Josee M. ;
Comans, Emile F. I. ;
Lammertsma, Adriaan A. ;
Paans, Anne M. ;
Willemsen, Antoon T. ;
Beyer, Thomas ;
Bockisch, Andreas ;
Schaefer-Prokop, Cornelia ;
Delbeke, Dominique ;
Baum, Richard P. ;
Chiti, Arturo ;
Krause, Bernd J. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (01) :181-200
[4]   The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials [J].
Boellaard, Ronald ;
Oyen, Wim J. G. ;
Hoekstra, Corneline J. ;
Hoekstra, Otto S. ;
Visser, Eric P. ;
Willemsen, Antoon T. ;
Arends, Bertjan ;
Verzijlbergen, Fred J. ;
Zijlstra, Josee ;
Paans, Anne M. ;
Comans, Emile F. I. ;
Pruim, Jan .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 (12) :2320-2333
[5]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[6]   Preoperative Staging of Lung Cancer with Combined PET-CT. [J].
Fischer, Barbara ;
Lassen, Ulrik ;
Mortensen, Jann ;
Larsen, Soren ;
Loft, Annika ;
Bertelsen, Anne ;
Ravn, Jesper ;
Clementsen, Paul ;
Hogholm, Asbjorn ;
Larsen, Klaus ;
Rasmussen, Torben ;
Keiding, Susanne ;
Dirksen, Asger ;
Gerke, Oke ;
Skov, Birgit ;
Steffensen, Ida ;
Hansen, Hanne ;
Vilmann, Peter ;
Jacobsen, Grete ;
Backer, Vibeke ;
Maltbaek, Niels ;
Pedersen, Jesper ;
Madsen, Henrik ;
Nielsen, Henrik ;
Hojgaard, Liselotte .
NEW ENGLAND JOURNAL OF MEDICINE, 2009, 361 (01) :32-39
[7]   Positron Emission Tomography in Staging Early Lung Cancer A Randomized Trial [J].
Maziak, Donna E. ;
Darling, Gail E. ;
Inculet, Richard I. ;
Gulenchyn, Karen Y. ;
Driedger, Albert A. ;
Ung, Yee C. ;
Miller, John D. ;
Gu, Chu-Shu ;
Cline, Kathryn J. ;
Evans, William K. ;
Levine, Mark N. .
ANNALS OF INTERNAL MEDICINE, 2009, 151 (04) :221-W48
[8]  
Shankar LK, 2006, J NUCL MED, V47, P1059
[9]   Effectiveness of positron emission tomography in the preoperative assessment of patients with suspected non-small-cell lung cancer: the PLUS multicentre randomised trial [J].
van Tinteren, H ;
Hoekstra, OS ;
Smit, EF ;
van den Bergh, JHAM ;
Schreurs, AJM ;
Stallaert, RALM ;
van Velthoven, PCM ;
Comans, EFI ;
Diepenhorst, FW ;
Verboom, P ;
van Mourik, JC ;
Postmus, PE ;
Boers, M ;
Teule, GJJ .
LANCET, 2002, 359 (9315) :1388-1392