Quantitative EEG in assessment of anaesthetic depth: Comparative study of methodology

被引:34
作者
Thomsen, CE [1 ]
Prior, PF [1 ]
机构
[1] UNIV LONDON ST BARTHOLOMEWS HOSP & MED COLL,DEPT CLIN NEUROPHYSIOL,LONDON EC1A 7BE,ENGLAND
关键词
monitoring; electroencephalography; anaesthesia; depth; memory; anaesthetics volatile; isoflurane; halothane;
D O I
10.1093/bja/77.2.172
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Methodology for assessment of depth of anaesthesia based on analysis of the electroencephalogram (EEG) is controversial. Techniques range from display of single measures,for example median value of the frequency spectrum, to dedicated pattern recognition systems based on measures of several EEG features. We have compared the performance of four techniques using tape-recorded data from 23 patients anaesthetized with either halothane or isoflurane using standardized regimens. The techniques were: (1) median frequency, (2) spectral edge frequency, (3) the cerebral function analysing monitor (CFAMI) and (4) a depth of anaesthesia monitor based on EEG pattern recognition (ADAM). Dose-response curves are presented for stepwise increases in stable end-tidal concentrations of each agent. Results indicated considerable inter-patient variability and the limitations of single EEG measures, particularly with deeper anaesthesia producing a burst suppression pattern in the EEG. Pattern recognition techniques reduced these difficulties and appeared to be promising over a wide range of anaesthetic levels.
引用
收藏
页码:172 / 178
页数:7
相关论文
共 22 条
[1]   AUTOMATIC ELECTROENCEPHALOGRAPHIC CONTROL OF GENERAL ANESTHESIA [J].
BICKFORD, RG .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1950, 2 (01) :93-96
[2]   A COMPARISON OF MEDIAN FREQUENCY, SPECTRAL EDGE FREQUENCY, A FREQUENCY BAND POWER RATIO, TOTAL POWER, AND DOMINANCE SHIFT IN THE DETERMINATION OF DEPTH OF ANESTHESIA [J].
DRUMMOND, JC ;
BRANN, CA ;
PERKINS, DE ;
WOLFE, DE .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1991, 35 (08) :693-699
[3]   EVALUATION OF THE ACTIONS OF GENERAL-ANESTHETICS IN THE HUMAN BRAIN [J].
JESSOP, J ;
JONES, JG .
GENERAL PHARMACOLOGY, 1992, 23 (06) :927-935
[4]   PERCEPTION AND MEMORY DURING GENERAL-ANESTHESIA [J].
JONES, JG .
BRITISH JOURNAL OF ANAESTHESIA, 1994, 73 (01) :31-37
[5]   INTRAOPERATIVE EEG PATTERNS - IMPLICATIONS FOR EEG MONITORING [J].
LEVY, WJ .
ANESTHESIOLOGY, 1984, 60 (05) :430-434
[6]   EFFECT OF EPOCH LENGTH ON POWER SPECTRUM ANALYSIS OF THE EEG [J].
LEVY, WJ .
ANESTHESIOLOGY, 1987, 66 (04) :489-495
[7]   AUTOMATED EEG PROCESSING FOR INTRA-OPERATIVE MONITORING - A COMPARISON OF TECHNIQUES [J].
LEVY, WJ ;
SHAPIRO, HM ;
MARUCHAK, G ;
MEATHE, E .
ANESTHESIOLOGY, 1980, 53 (03) :223-236
[8]   QUANTITATIVE EEG AND BRAIN-STEM AUDITORY EVOKED-POTENTIALS - COMPARISON OF ISOFLURANE WITH HALOTHANE USING THE CEREBRAL FUNCTION ANALYZING MONITOR [J].
LLOYDTHOMAS, AR ;
COLE, PV ;
PRIOR, PF .
BRITISH JOURNAL OF ANAESTHESIA, 1990, 65 (03) :306-312
[9]  
Lui WHD, 1991, ANAESTHESIA, V46, P435
[10]  
LUNN JN, 1987, CONSCIOUSNESS AWAREN