Background. Chronic haemodialysis (HD) may relieve some medical problems of terminal uraemia, but the life expectancy of patients is still significantly shortened, and there is a greatly increased morbidity. This includes pulmonary morbidity and chronic central nervous system (CNS) abnormalities. Previous studies have shown that a considerable amount of air microbubbles emanate within the blood lines of the dialysis device and pass the air detector without sounding an alarm. The aim of this study was to investigate whether microemboli can pass to the patient and whether they could be detected in the carotid artery. Methods. A total of 54 patients on chronic HD (16 with central dialysis catheter) were investigated with an ultrasound detector (Hatteland, Royken, Norway) for the presence of microemboli at the arteriovenous (AV) fistula/graft and at the common carotid artery before and during HD. Measurements were taken for 2 and 5 min, respectively. Non-parametric paired statistics were used (Wilcoxon). Results. The median number (range) and mean +/- SD of microembolic signals detected at the AV access site before commencing dialysis and during HD were 0 (0-3) and 0.2 +/- 0.5 versus 4 (0-85) and 13.5 +/- 20 (P = 0.000); at the carotid artery, 1 (0-14) and 1.7 +/- 2.9 versus 2 (0-36) and 3.5 +/- 5.8 (P = 0.008). Conclusions. The infused and returning fluid from HD devices contains air microbubbles that enter the patient without triggering any alarms. These small emboli pass the lung and may cause ischaemic lesions in organs supported by the arterial circuit, such as the brain.
引用
收藏
页码:2691 / 2695
页数:5
相关论文
共 46 条
[1]   Pulmonary hypertension in chronic dialysis patients with arteriovenous fistula: pathogenesis and therapeutic prospective [J].
Abassi, Zaid ;
Nakhoul, Farid ;
Khankin, Eliyahu ;
Reisner, Shimon A. ;
Yigla, Mordechai .
CURRENT OPINION IN NEPHROLOGY AND HYPERTENSION, 2006, 15 (04) :353-360
[2]   Pulmonary hypertension in patients with chronic renal failure - Role of parathyroid hormone and pulmonary artery calcifications [J].
Amin, M ;
Fawzy, A ;
Hamid, MA ;
Elhendy, A .
CHEST, 2003, 124 (06) :2093-2097
[3]   Pathophysiology and clinical implications of microbubbles during hemodialysis [J].
Barak, Michal ;
Nakhoul, Farid ;
Katz, Yeshayahu .
SEMINARS IN DIALYSIS, 2008, 21 (03) :232-238
[4]   EVIDENCE FOR PULMONARY MICROEMBOLIZATION DURING HEMODIALYSIS [J].
BISCHEL, MD ;
SCOLES, BG ;
MOHLER, JG .
CHEST, 1975, 67 (03) :335-337
[5]   Theoretical and experimental intravascular gas embolism absorption dynamics [J].
Branger, AB ;
Eckmann, DM .
JOURNAL OF APPLIED PHYSIOLOGY, 1999, 87 (04) :1287-1295
[6]  
Buoncristiani U, 1993, Perit Dial Int, V13 Suppl 2, pS202
[7]   LUNG AS A FILTER FOR MICROBUBBLES [J].
BUTLER, BD ;
HILLS, BA .
JOURNAL OF APPLIED PHYSIOLOGY, 1979, 47 (03) :537-543
[8]  
Coskun M, 1999, ACTA RADIOL, V40, P181
[9]   Reduction of circulating microemboli in the subclavian vein of patients undergoing haemodialysis using pre-filled instead of dry dialysers [J].
Droste, DW ;
Beyna, T ;
Frye, B ;
Schulte, V ;
Ringelstein, EB ;
Schaefer, RM .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2003, 18 (11) :2377-2381
[10]   Detection of microemboli in the subclavian vein of patients undergoing haemodialysis and haemodifiltration using pulsed Doppler ultrasound [J].
Droste, DW ;
Kühne, K ;
Schaefer, RM ;
Ringelstein, EB .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2002, 17 (03) :462-466