Automated methods for multiplexed pathogen detection

被引:50
作者
Straub, TM [1 ]
Dockendorff, BP [1 ]
Quiñonez-Díaz, MD [1 ]
Valdez, CO [1 ]
Shutthanandan, JI [1 ]
Tarasevich, BJ [1 ]
Grate, JW [1 ]
Bruckner-Lea, CJ [1 ]
机构
[1] Pacific NW Natl Lab, Interfacial Chem & Engn Grp, Richland, WA 99352 USA
关键词
analytical separations; automated methods; microarrays; pathogens; biodetection;
D O I
10.1016/j.mimet.2005.04.012
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Detection of pathogenic microorganisms in environmental samples is a difficult process. Concentration of the organisms of interest also co-concentrates inhibitors of many end-point detection methods, notably, nucleic acid methods. In addition, sensitive, highly multiplexed pathogen detection continues to be problematic. The primary function of the BEADS (Biodetection Enabling Analyte Delivery System) platform is the automated concentration and purification of target analytes from interfering substances, often present in these samples, via a renewable surface column. In one version of BEADS, automated immunomagnetic separation (IMS) is used to separate cells from their samples. Captured cells are transferred to a flow-through thermal cycler where PCR, using labeled primers, is performed. PCR products are then detected by hybridization to a DNA suspension array. In another version of BEADS, cell lysis is performed, and community RNA is purified and directly labeled. Multiplexed detection is accomplished by direct hybridization of the RNA to a planar microarray. The integrated IMS/PCR version of BEADS can successfully purify and amplify 10 E. coli O157:H7 cells from river water samples. Multiplexed PCR assays for the simultaneous detection of E. coli O157:H7, Salmonella, and Shigella on bead suspension arrays was demonstrated for the detection of as few as 100 cells for each organism. Results for the RNA version of BEADS are also showing promising results. Automation yields highly purified RNA, suitable for multiplexed detection on microarrays, with microarray detection specificity equivalent to PCR. Both versions of the BEADS platform show great promise for automated pathogen detection from environmental samples. Highly multiplexed pathogen detection using PCR continues to be problematic, but may be required for trace detection in large volume samples. The RNA approach solves the issues of highly multiplexed PCR and provides "live vs. dead" capabilities. However, sensitivity of the method will need to be improved for RNA analysis to replace PCR. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:303 / 316
页数:14
相关论文
共 44 条
[1]   The isotope array, a new tool that employs substrate-mediated labeling of rRNA for determination of microbial community structure and function [J].
Adamczyk, J ;
Hesselsoe, M ;
Iversen, N ;
Horn, M ;
Lehner, A ;
Nielsen, PH ;
Schloter, M ;
Roslev, P ;
Wagner, M .
APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 2003, 69 (11) :6875-6887
[2]   Separating microbes in the manner of molecules. 1. Capillary electrokinetic approaches [J].
Armstrong, DW ;
Schulte, G ;
Schneiderheinze, JM ;
Westenberg, DJ .
ANALYTICAL CHEMISTRY, 1999, 71 (24) :5465-5469
[3]   Portable system for microbial sample preparation and oligonucleotide microarray analysis [J].
Bavykin, SG ;
Akowski, JP ;
Zakhariev, VM ;
Barsky, VE ;
Perov, AN ;
Mirzabekov, AD .
APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 2001, 67 (02) :922-928
[4]   DETECTION OF COLIFORM BACTERIA IN WATER BY POLYMERASE CHAIN-REACTION AND GENE PROBES [J].
BEJ, AK ;
STEFFAN, RJ ;
DICESARE, J ;
HAFF, L ;
ATLAS, RM .
APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 1990, 56 (02) :307-314
[5]   A reusable flow-through polymerase chain reaction instrument for the continuous monitoring of infectious biological agents [J].
Belgrader, P ;
Elkin, CJ ;
Brown, SB ;
Nasarabadi, SN ;
Langlois, RG ;
Milanovich, FP ;
Colston, BW ;
Marshall, GD .
ANALYTICAL CHEMISTRY, 2003, 75 (14) :3446-3450
[6]  
Belgrader P, 1998, CLIN CHEM, V44, P2191
[7]  
Berry ED, 1999, J RAPID METH AUT MIC, V7, P7
[8]   Hydroxyapatite adherence as a means to concentrate bacteria [J].
Berry, ED ;
Siragusa, GR .
APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 1997, 63 (10) :4069-4074
[9]   Renewable microcolumns for automated DNA purification and flow-through amplification: from sediment samples through polymerase chain reaction [J].
Bruckner-Lea, CJ ;
Tsukuda, T ;
Dockendorff, B ;
Follansbee, JC ;
Kingsley, MT ;
Ocampo, C ;
Stults, JR ;
Chandler, DP .
ANALYTICA CHIMICA ACTA, 2002, 469 (01) :129-140
[10]   Integrated systems for DNA sample preparation and detection in environmental samples [J].
Bruckner-Lea, CJ ;
Anheier, NC ;
Holman, D ;
Tsukuda, T ;
Kingsley, MT ;
Brockman, FJ ;
Price, JM ;
Grate, JW ;
Chandler, DP .
BIOCHEMICAL AND BIOMOLECULAR SENSING, 2000, 4200 :74-81