Image degradation by glare in radiologic display devices

被引:13
作者
Badano, A
Flynn, MJ
机构
来源
IMAGE DISPLAY - MEDICAL IMAGING 1997 | 1997年 / 3031卷
关键词
flat panel display; emissive display; digital radiography; CRT; glare;
D O I
10.1117/12.273900
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
No electronic devices are currently available that can display digital radiographs without loss of visual information compared to traditional transilluminated film. Light scattering within the glass faceplate of cathode-ray tube (CRT) devices causes excessive glare that reduces image contrast. This glare, along with ambient light reflection has been recognized as a significant limitation for radiologic applications. Efforts to control the effect of glare and ambient light reflection in CRTs include the use of absorptive glass and thin film coatings. In the near future, hat panel displays (FPD) with thin emissive structures should provide very low glare, high performance devices. We have used an optical Monte Carlo simulation to evaluate the effect of glare on image quality for typical CRT and flat panel display devices. The trade-off between display brightness and image contrast is described. For CRT systems, achieving good glare ratio requires a reduction of brightness to 30-40 % of the maximum potential brightness. For FPD systems, similar glare performance can be achieved while maintaining 80 % of the maximum potential brightness.
引用
收藏
页码:222 / 231
页数:10