Low-bandwidth tele-ultrasound

被引:8
作者
Brebner, JA [1 ]
Ruddick-Bracken, H
Brebner, EM
Gilbert, FJ
Maclean, JR
Ritchie, LDR
Smith, P
Thompson, A
机构
[1] Univ Aberdeen, Dept Gen Practice & Primary Care, Aberdeen AB9 2AY, Scotland
[2] Univ Aberdeen, Dept Radiol, Aberdeen AB9 2AY, Scotland
[3] Med Coll Georgia, Dept Med, Augusta, GA 30912 USA
[4] Aberdeen Royal Hosp NHS Trust, Aberdeen Matern Hosp, Aberdeen, Scotland
[5] Grampian Healthcare NHS Trust, Dept Radiol, Aberdeen, Scotland
关键词
D O I
10.1258/1357633991932658
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
We examined the acceptability and diagnostic accuracy of dynamic ultrasound images transmitted at 128 kbit/s and 384 kbit/s. The gold standard was the direct recording of 200 ultrasound examinations on video-tape. The taped images were later transmitted at both 128 kbit/s and 384 kbit/s and recorded, resulting in three tapes for each case. Four observers viewed each tape individually. Ninety per cent of images transmitted at 384 kbit/s were rated as diagnostically acceptable compared with 32% of images transmitted at 128 kbit/s. Diagnostic agreement between tapes transmitted at 384 kbit/s and the gold standard was 85%, compared with 78% for 128 kbit/s transmissions. Observers were not satisfied with low-bandwidth transmission of ultrasound images despite adequate diagnostic accuracy. Dynamic ultrasound images transmitted at 384 kbit/s were viewed as both diagnostically acceptable and accurate.
引用
收藏
页码:75 / 76
页数:2