Hardware-related complications of deep brain stimulation: a ten year experience

被引:163
作者
Blomstedt, P [1 ]
Hariz, MI
机构
[1] Umea Univ Hosp, Dept Neurosurg, SE-90185 Umea, Sweden
[2] Inst Neurol, London WC1N 3BG, England
关键词
deep brain stimulation; stereotactic surgery; complications;
D O I
10.1007/s00701-005-0576-5
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective. To analyse the occurrence of hardware-related complications in patients with deep brain stimulation (DBS), over a long period of time. Method. All patients operated on with DBS at our institution between 1993 and 2002 were followed with respect to adverse events related to the implanted hardware. Results. One hundred and nineteen consecutive patients underwent 139 procedures with implantation of 161 electrodes. The minimum follow-up was 12 months. The follow-up time was 540 electrode-years. The rate of hardware-related complications per electrode-year was 4.3%. In total, 17 patients (15%) had 23 hardware-related complications. These included 8 electrode breakages, 4 electrode migrations, 2 stimulator migrations, 3 erosions, 2 erosions and infections, 2 infections and 2 cases of stimulator malfunction. The majority of these complications occurred during the first four years in our experience. Conclusions. DBS is a life-long therapy that requires a life-long follow-up. Increased experience and adaptation of surgical technique are the main determinants for avoidance of hardware-related complications.
引用
收藏
页码:1061 / 1064
页数:4
相关论文
共 16 条
  • [1] BENABID AL, 1987, APPL NEUROPHYSIOL, V50, P344
  • [2] Complications of deep brain stimulation surgery
    Beric, A
    Kelly, PJ
    Rezai, A
    Sterio, D
    Mogilner, A
    Zonenshayn, M
    Kopell, B
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2001, 77 (1-4) : 73 - 78
  • [3] Technical complication in deep brain stimulation
    Hamel, W
    Schrader, B
    Weinert, D
    Herzog, J
    Müller, D
    Deuschl, G
    Volkmann, J
    Mehdorn, HM
    [J]. ZENTRALBLATT FUR NEUROCHIRURGIE, 2002, 63 (03): : 124 - 127
  • [4] Hariz M, 1996, TXB STEREOTACTIC FUN, P87
  • [5] Hariz MI, 2001, MOVEMENT DISORD, V16, P166, DOI 10.1002/1531-8257(200101)16:1<166::AID-MDS1031>3.0.CO
  • [6] 2-S
  • [7] Comparison between stereotactic CT and MRI coordinates of pallidal and thalamic targets using the laitinen noninvasive stereoadapter
    Hirabayashi, H
    Hariz, MI
    Fagerlund, M
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1998, 71 (03) : 117 - 130
  • [8] Hardware-related problems of deep brain stimulation
    Joint, C
    Nandi, D
    Parkin, S
    Gregory, R
    Aziz, T
    [J]. MOVEMENT DISORDERS, 2002, 17 : S175 - S180
  • [9] Hardware-related complications after placement of thalamic deep brain stimulator systems
    Kondziolka, D
    Whiting, D
    Germanwala, A
    Oh, M
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2002, 79 (3-4) : 228 - 233
  • [10] LUNSFORD LD, 1996, TXB STEREOTACTIC FUN, P51