Configuration of the SCOTI device with different tracheal tubes

被引:4
作者
Nandwani, N
Caranza, R
Lin, ES
Raphael, JH
机构
[1] University Dept of Anaesthesia, Leicester Royal Infirmary
关键词
intubation; tracheal; equipment; sonomatic confirmation of tracheal intubation (SCOTI);
D O I
10.1111/j.1365-2044.1996.tb14960.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have evaluated the Sonomatic Confirmation of Tracheal Intubation device (SCOTI) by testing its ability to be correctly configured with a variety of tracheal tubes of differing internal diameter and length. The device only configured correctly for RAE tubes with internal diameter of 7.0 mm or greater and for armoured tubes of internal diameter 8.5 mm. For conventional tubes of varying internal diameter cut to different lengths, configuration was only successful with certain dimensions. the inability to configure the device correctly with all types and lengths of tracheal tubes limits its usefulness as a indicator of tracheal intubation.
引用
收藏
页码:932 / 934
页数:3
相关论文
共 2 条
[1]   SCOTI - A NEW DEVICE FOR IDENTIFICATION OF TRACHEAL INTUBATION [J].
MURRAY, D ;
WARD, ME ;
SEAR, JW .
ANAESTHESIA, 1995, 50 (12) :1062-1064
[2]  
*PENL UK LTD, 1995, SCOTI US MAN