Left ventricular assist device malfunction - An approach to diagnosis by echocardiography

被引:58
作者
Horton, SC
Khodaverdian, R
Chatelain, P
McIntosh, ML
Horne, BD
Muhlestein, JB
Long, JW
机构
[1] Univ Utah, Sch Med, Div Utah Artificial Heart Program, Salt Lake City, UT USA
[2] Univ Utah, Sch Med, LDS Hosp, Dept Cardiol, Salt Lake City, UT USA
关键词
D O I
10.1016/j.jacc.2005.01.037
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A protocol using transthoracic echocardiography was designed to diagnose the common malfunctions of patients on chronic support with a left ventricular assist device (LVAD). Mechanical circulatory support, primarily with a LVAD, is increasingly used for treatment of advanced heart failure as a bridge to transplant and for long-term treatment of heart failure. The LVAD dysfunction is a recognized complication. To date, no studies have defined the role of transthoracic echocardiography in evaluating long-term mechanical complications of chronic LVAD support. Transthoracic echocardiography was used in a protocol designed to detect the common types of mechanical malfunction. Patients were followed up with serial echocardiograms, and clinical validations were made with findings from a catheter-based protocol and inspection at the time of cardiac transplant or corrective surgery. Thirty-two patients with 44 LVADs were followed up during a four-year period using this protocol that correctly identified 11 patients with inflow valve regurgitation, 2 with intermittent inflow conduit obstruction, 1 with severe kinking of the outflow graft, and 9 with new insufficiency of the native aortic valve. As LVAD use for end-stage heart failure becomes widespread, and durations of support are extended, dysfunction will be increasingly prevalent. Transthoracic echocardiography provides a practical method to accurately identify the causes of mechanical dysfunction with patients on chronic LVAD support. (c) 2005 by the American College of Cardiology Foundation.
引用
收藏
页码:1435 / 1440
页数:6
相关论文
共 10 条
[1]   Echocardiographic evaluation of patients with a left ventricular device [J].
Akosah, KO ;
Song, A ;
Guerraty, A ;
Mohanty, P ;
Paulsen, W .
ASAIO JOURNAL, 1998, 44 (05) :M624-M627
[2]   Effects of left ventricular assist devices on outcomes in patients undergoing heart transplantation [J].
Bank, AJ ;
Mir, SH ;
Nguyen, DQ ;
Bolman, RM ;
Shumway, SJ ;
Miller, LW ;
Kaiser, DR ;
Ormaza, SM ;
Park, SJ .
ANNALS OF THORACIC SURGERY, 2000, 69 (05) :1369-1374
[3]   Acquired commissural fusion of aortic valves in patients with left ventricular assist devices [J].
Connelly, JH ;
Abrams, J ;
Klima, T ;
Vaughn, WK ;
Frazier, OH .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2003, 22 (12) :1291-1295
[4]   Evaluation of a patient with left ventricular assist device dysfunction [J].
Ferns, J ;
Dowling, R ;
Bhat, G .
ASAIO JOURNAL, 2001, 47 (06) :696-698
[5]   IMPROVED MORTALITY AND REHABILITATION OF TRANSPLANT CANDIDATES TREATED WITH A LONG-TERM IMPLANTABLE LEFT-VENTRICULAR ASSIST SYSTEM [J].
FRAZIER, OH ;
ROSE, EA ;
MCCARTHY, P ;
BURTON, NA ;
TECTOR, A ;
LEVIN, H ;
KAYNE, HL ;
POIRIER, VL ;
DASSE, KA .
ANNALS OF SURGERY, 1995, 222 (03) :327-338
[6]   Left ventricular assist device malfunction: A systematic approach to diagnosis [J].
Horton, SC ;
Khodaverdian, R ;
Powers, A ;
Revenaugh, J ;
Renlund, DG ;
Moore, SA ;
Rasmusson, B ;
Nelson, KE ;
Long, JW .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (09) :1574-1583
[7]  
POIRIER V, 2001, J CONGESTIVE HEART F, V2, P23
[8]   Partial aortic valve fusion induced by left ventricular assist device [J].
Rose, AG ;
Park, SJ ;
Bank, AJ ;
Miller, LW .
ANNALS OF THORACIC SURGERY, 2000, 70 (04) :1270-1274
[9]   Long-term use of a left ventricular assist device for end-stage heart failure [J].
Rose, EA ;
Gelijns, AC ;
Moskowitz, AJ ;
Heitjan, DF ;
Stevenson, LW ;
Dembitsky, W ;
Long, JW ;
Ascheim, DD ;
Tierney, AR ;
Levitan, RG ;
Watson, JT ;
Meier, P .
NEW ENGLAND JOURNAL OF MEDICINE, 2001, 345 (20) :1435-1443
[10]   Clinical utility of echocardiography in the management of implantable ventricular assist devices [J].
Scalia, GM ;
McCarthy, PM ;
Savage, RM ;
Smedira, NG ;
Thomas, JD .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2000, 13 (08) :754-763