A low Power ECG Signal Processor for Ambulatory Arrhythmia Monitoring System

被引:27
作者
Kim, Hyejung [1 ]
Yazicioglu, Refet Firat [1 ]
Torfs, Tom [1 ]
Merken, Patrick [1 ]
Yoo, Hoi-Jun [2 ]
Van Hoof, Chris [1 ]
机构
[1] IMEC, Kapeldreef 75, B-3001 Louvain, Belgium
[2] Korea Adv Inst Sci & Technol, Daejeon, South Korea
来源
2010 SYMPOSIUM ON VLSI CIRCUITS, DIGEST OF TECHNICAL PAPERS | 2010年
关键词
biomedical signal processing; ECG; processor;
D O I
10.1109/VLSIC.2010.5560253
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An ECG signal processor ( ESP) is proposed for ambulatory arrhythmia monitoring systems. The ESP consists of three heterogeneous processors and performs filtering, data compression, ECG classification, and encryption. A data reduction scheme, consisting of skeleton and Huffman coding, are employed to reduce the on-chip memory capacity and memory access power. Clock gating and voltage scaling are also applied to reduce the power consumption. The ESP consumes 1.26-mu W at 0.7V, while providing real time signal processing.
引用
收藏
页码:19 / +
页数:2
相关论文
共 5 条
[1]  
JOCKE SC, 2009, IEEE P VLSI S JUN
[2]  
KIM H, 2009, IEEE P EMBC SEP
[3]   ECG Signal Compression and Classification Algorithm With Quad Level Vector for ECG Holter System [J].
Kim, Hyejung ;
Yazicioglu, Refet Firat ;
Merken, Patrick ;
Van Hoof, Chris ;
Yoo, Hoi-Jun .
IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2010, 14 (01) :93-100
[4]  
Reddy DC, 2005, BIOMEDICAL SIGNAL PR
[5]  
YAZICIOGLU RF, 2009, IEEE P EMBC SEP