On the usefulness of the mechanical index displayed on clinical ultrasound scanners for predicting contrast microbubble destruction

被引:27
作者
Forsberg, F [1 ]
Shi, WT [1 ]
Merritt, CRB [1 ]
Dai, Q [1 ]
Solcova, M [1 ]
Goldberg, BB [1 ]
机构
[1] Thomas Jefferson Univ, Dept Radiol, Philadelphia, PA 19107 USA
关键词
acoustic field conditions; mechanical index; microbubble destruction; sonographic contrast media;
D O I
10.7863/jum.2005.24.4.443
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective. The purpose of this study was to evaluate the mechanical index (MI) displayed on clinical ultrasound scanners as a predictor of exposure conditions related to the destruction of sonographic microbubble contrast agents. Methods. Sonazoid (GE Healthcare, Oslo, Norway) and Optison (GE Healthcare, Princeton, NJ) microbubbles were injected into a tissue-mimicking flow phantom. Gray scale imaging was performed with 4 different scanners and 3 different transducers (3.5 MHz curved linear, 2.5 MHz convex, and 7.5 MHz linear array), and the MI displayed by the scanner was varied from 0.2 to 1.5 by changing the system output power. All other scanning parameters were kept constant. Downstream changes in echogenicity were monitored with a PowerVision 7000 scanner (Toshiba America Medical Systems, Tustin, CA) as an indirect measure of bubble destruction. Video intensity changes within the flow tube were determined as a function of MI for the different scanner/transducer combinations, and the best linear fit was determined. Results. At a displayed MI of 0.7, different scanner/transducer combinations exhibited a range in video intensity from +16% to -3% of baseline for Sonazoid and from +8% to -71% for Optison. At an MI of 0.3, reductions in video intensity of up to 32% were produced. These results indicate a wide range in bubble destruction at identical MI values. Likewise, regression analysis found no linear fits for all scanner/transducer combinations (r(2) < 0.046). Conclusions. The MI displayed on clinical ultrasound scanners does not predict the degree of microbubble destruction and should not be used by itself to define exposure conditions for destruction of microbubble contrast agents.
引用
收藏
页码:443 / 450
页数:8
相关论文
共 38 条
[1]  
*AM I ULTR MED, 2002, BIOEFF DIAGN ULTR GA
[2]  
American Institute of Ultrasound in Medicine and National Electrical Manufacturers Association, 1992, STAND REAL TIM DISPL
[3]   GAUGING THE LIKELIHOOD OF CAVITATION FROM SHORT-PULSE, LOW-DUTY CYCLE DIAGNOSTIC ULTRASOUND [J].
APFEL, RE ;
HOLLAND, CK .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (02) :179-185
[4]   Science, medicine, and the future - Microbubble contrast agents: a new era in ultrasound [J].
Blomley, MJK ;
Cooke, JC ;
Unger, EC ;
Monaghan, MJ ;
Cosgrove, DO .
BMJ-BRITISH MEDICAL JOURNAL, 2001, 322 (7296) :1222-1225
[5]   BACKGROUND SUBTRACTION FROM TIME-INTENSITY CURVES IN VIDEODENSITOMETRY - A PITFALL IN FLOW ASSESSMENT USING CONTRAST ECHOCARDIOGRAPHY [J].
BOS, LJ ;
PIEK, JJ ;
SPAAN, JAE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (09) :1211-1218
[6]   Mechanisms of contrast agent destruction [J].
Chomas, JE ;
Dayton, P ;
Allen, J ;
Morgan, K ;
Ferrara, KW .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2001, 48 (01) :232-248
[7]   Focal hepatic masses: Enhancement patterns with SHU 508A and pulse-inversion US [J].
Dill-Macky, MJ ;
Burns, PN ;
Khalili, K ;
Wilson, SR .
RADIOLOGY, 2002, 222 (01) :95-102
[8]   Hepatic tumor detection: MR Imaging and conventional US versus pulse-inversion harmonic US of NC100100 during its reticuloendothelial system-specific phase [J].
Forsberg, F ;
Piccoli, CW ;
Liu, JB ;
Rawool, NM ;
Merton, DA ;
Mitchell, DG ;
Goldberg, BB .
RADIOLOGY, 2002, 222 (03) :824-829
[9]   On the feasibility of real time, in vivo harmonic imaging with proteinaceous microspheres [J].
Forsberg, F ;
Goldberg, BB ;
Liu, JB ;
Merton, DA ;
Rawool, NM .
JOURNAL OF ULTRASOUND IN MEDICINE, 1996, 15 (12) :853-860
[10]  
Forsberg F, 2000, J ULTRAS MED, V19, P557