Structured approach to achieving high accuracy measurements with isotope dilution inductively coupled plasma mass spectrometry

被引:59
作者
Catterick, T [1 ]
Fairman, B [1 ]
Harrington, C [1 ]
机构
[1] LGC Teddington Ltd, Teddington TW11 0LY, Middx, England
关键词
isotope dilution; inductively coupled plasma mass spectrometry; accuracy; precision; error; matched isotope ratio; certified reference materials;
D O I
10.1039/a801233d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study investigated a method for simplifying the implementation of ID-ICP-MS, to achieve high accuracy measurements. The method developed is an adaptation of an earlier methodology that utilised an iterative 'matching' procedure. While retaining the many advantages of this approach, ne have removed the iterative component and simplified the 'matching' step between the spiked mass bias calibrant and the spiked sample. Based on a conventional analysis, an approximate 'match' (usually within 5%), was made between the spiked reference standard and the spiked sample. This provides the basis for calculating the final result using the normal IDMS equation, while retaining many of the benefits of the full 'matching' procedure, These benefits negate many sources of error, such as mass bias, detector dead time and characterisation of the spike material, Many common errors and necessary corrections are negated or eliminated when using the new procedure. Examples are given of the method as applied to the analysis of certified reference materials, such as waters and plastics as well as blind trial data. The accuracy achieved using this procedure, on blind trial solutions, is typically to within 1% (relative to concentration) at the 95% level of confidence.
引用
收藏
页码:1009 / 1013
页数:5
相关论文
共 15 条
[1]  
[Anonymous], 1988, INORGANIC MASS SPECT
[2]   REFERENCE MATERIALS BY ISOTOPE RATIO MASS-SPECTROMETRY [J].
DEBIEVRE, P ;
DELAETER, JR ;
PEISER, HS ;
REED, WP .
MASS SPECTROMETRY REVIEWS, 1993, 12 (03) :143-172
[3]   PRECISION MASS SPECTROMETRIC ISOTOPE DILUTION ANALYSIS [J].
DEBIEVRE, PJ ;
DEBUS, GH .
NUCLEAR INSTRUMENTS & METHODS, 1965, 32 (02) :224-&
[4]  
FASSETT JD, 1989, ANAL CHEM, V61, P643
[5]   REDUCTION OF SYSTEMATIC-ERRORS IN QUANTITATIVE-ANALYSIS BY ISOTOPE-DILUTION MASS-SPECTROMETRY (IDMS) - AN ITERATIVE METHOD [J].
HENRION, A .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1994, 350 (12) :657-658
[6]   ISOTOPE-DILUTION MASS-SPECTROMETRY OF INORGANIC AND ORGANIC-SUBSTANCES [J].
HEUMANN, KG .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1986, 325 (08) :661-666
[7]  
HEUMANN KG, 1980, ADV MASS SPECTROM, V8, P313
[8]  
INGRAHM MG, 1950, PHYS REV, V80, P916
[9]   Establishing an SI-traceable copper concentration in the candidate reference material MURST ISS A1 Antarctic sediment using isotope dilution applied as a primary method of measurement [J].
Papadakis, I ;
Taylor, PDP ;
DeBievre, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (08) :791-796
[10]   A MASS SPECTROMETRIC INVESTIGATION OF BRANCHING IN CU-64, BR-80, BR-82 AND I-128 [J].
REYNOLDS, JH .
PHYSICAL REVIEW, 1950, 79 (05) :789-794