Electrochemical and surface characterization of a nickel-titanium alloy

被引:376
作者
Wever, DJ
Veldhuizen, AG
de Vries, J
Busscher, HJ
Uges, DRA
van Horn, JR
机构
[1] Univ Groningen Hosp, Dept Orthopaed, NL-9700 RB Groningen, Netherlands
[2] Univ Groningen, Mat Tech Lab, NL-9700 RB Groningen, Netherlands
[3] Univ Groningen Hosp, Dept Pharm, Groningen, Netherlands
关键词
NiTi; memory metal; corrosion; XPS; nickel release;
D O I
10.1016/S0142-9612(97)00210-X
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For clinical implantation purposes of shape memory metals the nearly equiatomic nickel-titanium (NiTi) alloy is generally used. In this study, the corrosion properties and surface characteristics of this alloy were investigated and compared with two reference controls, AISI 316 LVM stainless steel and Ti6Al4V. The anodic polarization curves, performed in Hanks' solution at 37 degrees C, demonstrated a passive behaviour for the NiTi alloy. A more pronounced difference between the corrosion and breakdown potential, i.e. a better resistance to chemical breakdown of passivity was found for the NiTi alloy compared to AISI 316 LVM. X-ray electron spectroscopy (XPS) and scanning electron microscopy (SEM) were undertaken to study the elemental composition and structure of the surface films prior to, and after immersion in Hanks' solution. The passive film on the NiTi alloy consists of a mainly TiO2-based oxide with minimal amounts of nickel in the outermost surface layers. After immersion in Hanks' solution the growth of a calcium-phosphate layer was observed. The passive diffusion of nickel from the NiTi alloy, measured by atomic absorption spectrophotometry reduced significantly in time from an initial release rate of 14.5 x 10(-7) mu g cm(-2) s(-1) to a nickel release that could not detect anymore after 10 days. It is suggested that the good corrosion properties of the NiTi alloy and the related promising biological response, as reported in literature, may be ascribed to the presence of mainly a TiO2-based surface layer and its specific properties, including the formation of a calcium-phosphate layer after exposure to a bioenvironment. (C) 1998 Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:761 / 769
页数:9
相关论文
共 34 条
[1]   THE SURFACE-ANALYSIS OF IMPLANT MATERIALS .1. THE SURFACE-COMPOSITION OF A TITANIUM DENTAL IMPLANT MATERIAL [J].
AMEEN, AP ;
SHORT, RD ;
JOHNS, R ;
SCHWACH, G .
CLINICAL ORAL IMPLANTS RESEARCH, 1993, 4 (03) :144-150
[3]   PREPARATION AND SURFACE SPECTROSCOPIC CHARACTERIZATION OF OXIDE-FILMS ON TI6A14V [J].
ASK, M ;
LAUSMAA, J ;
KASEMO, B .
APPLIED SURFACE SCIENCE, 1989, 35 (03) :283-301
[4]  
*ASTM, 1993, ANN BOOK ASTM STAND, P19
[5]  
*ASTM, 1994, ANN BOOK ASTM STAND, P63
[6]  
*ASTM, 1993, ANN BOOK ASTM STAND, P22
[7]  
BENSMANN G, 1982, TECH MITT KRUPP FORS, V40, P123
[8]   SURFACE MODIFICATION OF TITANIUM-ALLOY IMPLANTS [J].
BROWNE, M ;
GREGSON, PJ .
BIOMATERIALS, 1994, 15 (11) :894-898
[9]  
Buehler WJ., 1968, Ocean Engineering, V1, P105, DOI [DOI 10.1016/0029-8018(68)90019-X, 10.1016/0029-8018(68)90019-X, DOI 10.1016/0029]
[10]   NITRIC-ACID PASSIVATION OF TI6AL4V REDUCES THICKNESS OF SURFACE OXIDE LAYER AND INCREASES TRACE-ELEMENT RELEASE [J].
CALLEN, BW ;
LOWENBERG, BF ;
LUGOWSKI, S ;
SODHI, RNS ;
DAVIES, JE .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (03) :279-290