Comparison between chest digital tomosynthesis and CT as a screening method to detect artificial pulmonary nodules: a phantom study

被引:26
作者
Gomi, T. [1 ]
Nakajima, M. [2 ]
Fujiwara, H. [2 ]
Takeda, T.
Saito, K.
Umeda, T.
Sakaguchi, K.
机构
[1] Kitasato Univ, Sch Allied Hlth Sci, Minami Ku, Sagamihara, Kanagawa 2520373, Japan
[2] Dokkyo Med Univ Hosp, Dept Radiol, Koshigaya, Saitama, Japan
关键词
HIGH-RESOLUTION CT; LUNG-CANCER; PATHOLOGICAL CORRELATION; CLINICAL-EXPERIENCE; COMPUTED-TOMOGRAPHY; HUMAN OBSERVER; RADIOGRAPHS;
D O I
10.1259/bjr/12643098
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Objectives: The objective of this study was to evaluate the imaging capabilities of chest digital tomosynthesis (DT) as a screening method for the detection of artificial pulmonary nodules, and to compare its efficiency with that of CT. Methods: DT and CT were used to detect artificial pulmonary nodules (5 mm and 8 mm in diameter, ground-glass opacities) placed in a chest phantom. Using a three-dimensional filtered back-projection algorithm at acquisition angles of 8 degrees, 20 degrees, 30 degrees and 40 degrees, DT images of the desired layer thicknesses were reconstructed from the image data acquired during a single tomographic scan. Both standard and sharp CT reconstruction kernels were used, and the detectability index (DI) valves computed for both the DT scan acquisition angles and CT reconstruction kernel types were considered. For the observer study, we examined 50 samples of artificial pulmonary nodules using both DT and CT imaging. On the basis of evaluations made by five thoracic radiologists, a jackknife free-response receiver operating characteristic (JAFROC) study was performed to compare and assess the differences in detection accuracy between CT and DT imaging. Results: For each increased acquisition angle, DI obtained by DT imaging was similar to that obtained by CT imaging. The difference in the observer-averaged JAFROC figure of merit for the five readings was 0.0363 (95% confidence interval: -0.18, 0.26; F=0.101; p=0.75). Conclusion: With the advantages of a decreased radiation dose and the practical accessibility of examination, DT may be a useful alternative to CT for the detection of artificial pulmonary nodules.
引用
收藏
页码:E622 / E629
页数:8
相关论文
共 29 条
[1]
ELECTROFLUOROPLANIGRAPHY [J].
BAILY, NA ;
LASSER, EC ;
CREPEAU, RL .
RADIOLOGY, 1973, 107 (03) :669-671
[2]
Analysis of location specific observer performance data: Validated extensions of the jackknife free-response (JAFROC) method [J].
Chakraborty, Dev P. .
ACADEMIC RADIOLOGY, 2006, 13 (10) :1187-1193
[3]
Chakraborty DP, JAFROC SOFTWARE 2 3A
[4]
Digital tomosynthesis of the chest for lung nodule detection: Interim sensitivity results from an ongoing NIH-sponsored trial [J].
Dobbins, James T. ;
McAdams, H. Page ;
Song, Jae-Woo ;
Li, Christina M. ;
Godfrey, Devon J. ;
DeLong, David M. ;
Paik, Sang-Hyun ;
Martinez-Jimenez, Santiago .
MEDICAL PHYSICS, 2008, 35 (06) :2554-2557
[5]
Digital x-ray tomosynthesis: current state of the art and clinical potential [J].
Dobbins, JT ;
Godfrey, DJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (19) :R65-R106
[6]
Clinical potential of digital linear tomosynthesis imaging of total joint arthroplasty [J].
Gomi, Tsutomu ;
Hirano, Hiroshi .
JOURNAL OF DIGITAL IMAGING, 2008, 21 (03) :312-322
[8]
ImPACT, IMPACTS CT DOS TOOL
[9]
Internal Commission on Radiation Units and Measurements ( ICRU), 1996, 54 ICRU MEAS
[10]
Overview of two years of clinical experience of chest tomosynthesis at Sahlgrenska University Hospital [J].
Johnsson, A. A. ;
Vikgren, J. ;
Svalkvist, A. ;
Zachrisson, S. ;
Flinck, A. ;
Boijsen, M. ;
Kheddache, S. ;
Mansson, L. G. ;
Bath, M. .