Preliminary results after application of absorbable metal stents in patients with critical limb ischemia

被引:223
作者
Peeters, P
Bosiers, M
Verbist, J
Deloose, K
Heublein, B
机构
[1] Imelda Hosp, Dept Cardiovasc & Thorac Surg, B-2820 Bonheiden, Belgium
[2] AZ St Blasius, Dept Vasc Surg, Dendermonde, Belgium
关键词
infrapopliteal artery; critical limb ischemia; absorbable stents; angioplasty; restenosis;
D O I
10.1583/04-1349R.1
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To report the preliminary 3-month follow-up of a recently developed absorbable metal stent (AMS) for treatment of infrapopliteal lesions in patients with critical limb ischemia (CLI). Methods: Between December 2003 and January 2004, 20 patients (10 men; mean age 76 years, range 59-96) with symptomatic critical limb ischemia (CLI) due to high-grade (80% to 100%) infrapopliteal stenoses received 1 or 2 AMS devices for suboptimal angioplasty. Clinical examination and color-flow duplex imaging were performed in all patients at discharge and at 1 and 3 months. Results: Angiographic procedural success was achieved in all 20 patients. One patient died (non-procedure-related) during the periprocedural period. Three months after the procedure, primary clinical patency was 89.5% (17/19). No major or minor amputation was necessary in any of the patients, yielding a limb salvage rate of 100%. The average improvement in Rutherford class was 2.3 at the 3-month assessment. A comparison of postprocedural and 1-month color-flow duplex ultrasound and magnetic resonance images clearly indicated the ongoing absorption process of the AMS. Conclusions: After 3 months, the primary clinical patency and limb salvage rates suggest a potentially promising performance of these AMS devices in the treatment of below-knee lesions in CLI patients.
引用
收藏
页码:1 / 5
页数:5
相关论文
共 11 条
[1]  
Balmer H, 2002, J ENDOVASC THER, V9, P403, DOI 10.1583/1545-1550(2002)009<0403:BAICCL>2.0.CO
[2]  
2
[3]   Biocorrosion of magnesium alloys: a new principle in cardiovascular implant technology? [J].
Heublein, B ;
Rohde, R ;
Kaese, V ;
Niemeyer, M ;
Hartung, W ;
Haverich, A .
HEART, 2003, 89 (06) :651-656
[4]   Pathological analyses of long-term intracoronary Palmaz-Schatz stenting - Is its efficacy permanent [J].
Inoue, K ;
Abe, K ;
Ando, K ;
Shirai, S ;
Nishiyama, K ;
Nakanishi, M ;
Yamada, T ;
Sakai, K ;
Nakagawa, Y ;
Hamasaki, N ;
Kimura, T ;
Nobuyoshi, M ;
Miyamoto, TA .
CARDIOVASCULAR PATHOLOGY, 2004, 13 (02) :109-115
[5]  
Migliavacca F, 2004, BIOMECH MODEL MECHAN, V2, P205, DOI [10.1007/s10237-004-0039-6, 10.1007/S10237-004-0039-6]
[6]   Subintimal angioplasty of infrapopliteal occlusions in critically ischaemic limbs [J].
Nydahl, S ;
Hartshorne, T ;
Bell, PRF ;
Bolia, A ;
London, NJM .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1997, 14 (03) :212-216
[7]   A novel approach to temporary stenting:: degradable cardiovascular stents produced from corrodible metal -: results 6-18 months after implantation into New Zealand white rabbits [J].
Peuster, M ;
Wohlsein, P ;
Brügmann, M ;
Ehlerding, M ;
Seidler, K ;
Fink, C ;
Brauer, H ;
Fischer, A ;
Hausdorf, G .
HEART, 2001, 86 (05) :563-569
[8]  
Rastogi Sachin, 2004, Tech Vasc Interv Radiol, V7, P33, DOI 10.1053/j.tvir.2004.01.001
[9]  
Tamai H, 2000, CIRCULATION, V102, P399
[10]  
Tsuji Takafumi, 2003, Int J Cardiovasc Intervent, V5, P13