An arrhythmia detector and heart rate estimator for overnight polysomnography studies

被引:21
作者
Bartolo, A
Clymer, BD
Burgess, RC
Turnbull, JP
Golish, JA
Perry, MC
机构
[1] Ohio State Univ, Ctr Biomed Engn, Columbus, OH 43210 USA
[2] Cleveland Clin Fdn, Dept Neurol, Cleveland, OH 44195 USA
[3] Ohio State Univ, Dept Elect Engn, Columbus, OH 43210 USA
[4] Cleveland Clin Fdn, Dept Pulm & Crit Care Med, Cleveland, OH 44195 USA
关键词
arrhythmia detection; automatic ECG analysis; heart rate estimation; MIT-BIH arrhythmia database; polysomnography; QRS detection; signal quality monitoring; sleep; weighted cross-correlation coefficient;
D O I
10.1109/10.918590
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present an algorithm for automatic on-line analysis of the electrocardiography (ECG) channel acquired during overnight polysomnography (PSG) studies. The system is independent of ECG morphology, requires no manual initialization, and operates automatically throughout the night. It highlights likely occurrences of arrhythmias and intervals of bad signal quality while outputting a continual estimate of heart rate. Algorithm performance is validated against standard ECG databases and PSG data. Results demonstrate a minimal false negative rate and a low false positive rate for arrhythmia detection, and robustness over a wide range of noise contamination.
引用
收藏
页码:513 / 521
页数:9
相关论文
共 12 条
[1]  
BRADLEY TD, 1994, SLEEP BREATHING, pCH27
[2]   ADAPTIVE REAL-TIME WAVELET DETECTION [J].
COHEN, A ;
LANDSBERG, D .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (06) :332-340
[3]   AN EFFICIENT ALGORITHM FOR WAVEFORM ANALYSIS USING THE CORRELATION-COEFFICIENT [J].
COLLINS, SM ;
ARZBAECHER, RC .
COMPUTERS AND BIOMEDICAL RESEARCH, 1981, 14 (04) :381-389
[4]  
DOBBS S E, 1984, Journal of Clinical Engineering, V9, P197
[5]  
GUILLEMINAULT C, 1984, LANCET, V1, P126
[6]  
LINDECRANTZ KG, 1988, J BIOMED ENG, V10, P208
[7]  
NEIDERMEYER E, 1993, ELECTROENCEPHALOGRAP, pCH46
[8]  
PATEL JM, 1990, ARRHYTHMIAS DETECTIO
[9]   CHANGES IN PULSE TRANSIT-TIME AND PULSE-RATE AS MARKERS OF AROUSAL FROM SLEEP IN NORMAL SUBJECTS [J].
PITSON, D ;
CHHINA, N ;
KNIJN, S ;
VANHERWAADEN, M ;
STRADLING, J .
CLINICAL SCIENCE, 1994, 87 (02) :269-273
[10]  
STRADLING JR, 1993, HDB SLEEP RELATED BR, P100