A novel target-field method for magnetic resonance shim coils: III. Shielded zonal and tesseral coils

被引:49
作者
Forbes, LK
Crozier, S
机构
[1] Univ Tasmania, Sch Math & Phys, Hobart, Tas 7001, Australia
[2] Univ Queensland, Sch Informat Technol & Elect Engn, Brisbane, Qld 4072, Australia
关键词
D O I
10.1088/0022-3727/36/2/302
中图分类号
O59 [应用物理学];
学科分类号
摘要
This paper continues the development of a new approach for the design of shim and gradient coils, used in magnetic resonance imaging (MRI) applications. A cylindrical primary coil of radius a and length 2L is placed inside a co-axial shield cylinder of radius b. An active shielding strategy is used to create a desired target field at an arbitrarily specified (cylindrical) location within the primary coil, and to annul the field at a certain radius outside the shield. The form of the interior target field may be chosen arbitrarily by the designer, although zonal and tesseral harmonics are typically used in MRI applications. The method presented here designs coil windings on both the primary and shielding cylinders, to produce fields that conform to the specified interior target field and the annulled field exterior to the shield. An additional feature of the method presented here is that the target field inside the primary coil is matched at two different radii, to improve overall accuracy. The method is illustrated by designing several shielded shim coils, for creating higher order tesseral fields located asymmetrically within the coil. The simpler case of pure zonal fields is discussed separately and applied to the design of some higher order shielded coils.
引用
收藏
页码:68 / 80
页数:13
相关论文
共 23 条
[1]  
[Anonymous], 1985, COMPUTATIONAL METHOD
[2]   GRADIENT COIL DESIGN USING ACTIVE MAGNETIC SCREENING [J].
BOWTELL, R ;
MANSFIELD, P .
MAGNETIC RESONANCE IN MEDICINE, 1991, 17 (01) :15-21
[3]   Determining complicated winding patterns for shim coils using stream functions and the target-field method [J].
Brideson, MA ;
Forbes, LK ;
Crozier, S .
CONCEPTS IN MAGNETIC RESONANCE, 2002, 14 (01) :9-18
[4]   DESIGN AND EVALUATION OF SHIELDED GRADIENT COILS [J].
CARLSON, JW ;
DERBY, KA ;
HAWRYSZKO, KC ;
WEIDEMAN, M .
MAGNETIC RESONANCE IN MEDICINE, 1992, 26 (02) :191-206
[5]   Compact MRI magnet design by stochastic optimization [J].
Crozier, S ;
Doddrell, DM .
JOURNAL OF MAGNETIC RESONANCE, 1997, 127 (02) :233-237
[6]   GRADIENT-COIL DESIGN BY SIMULATED ANNEALING [J].
CROZIER, S ;
DODDRELL, DM .
JOURNAL OF MAGNETIC RESONANCE SERIES A, 1993, 103 (03) :354-357
[7]  
CROZIER S, 2002, IEEE T APPL SUPERCON
[8]   A novel target-field method for finite-length magnetic resonance shim coils: I. Zonal shims [J].
Forbes, LK ;
Crozier, S .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2001, 34 (24) :3447-3455
[9]   A novel target-field method for finite-length magnetic resonance shim coils: II. Tesseral shims [J].
Forbes, LK ;
Crozier, S .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2002, 35 (09) :839-849
[10]   Asymmetric zonal shim coils for magnetic resonance applications [J].
Forbes, LK ;
Crozier, S .
MEDICAL PHYSICS, 2001, 28 (08) :1644-1651