How precise is navigation for minimally invasive surgery?

被引:11
作者
Broers, H. [1 ]
Jansing, N. [1 ]
机构
[1] AXIOS 3DR Serv GmbH, D-26121 Oldenburg, Germany
关键词
Navigation System; Marker Ball; Invasive Operation Technique; Navigate Measurement; Statistical Experiment Method;
D O I
10.1007/s00264-007-0431-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Infra-red measurement systems are an established facet of operative procedures using surgical navigation, yet questions remain regarding their accuracy and everyday suitability. The importance of these systems has grown with the increasing popularity of minimally invasive procedures, which are associated with a reduction in operative field visibility. Even though these measurement systems are reliable and accurate, residual errors are still present and impact the overall success rate of surgical navigation. We performed empirical experiments to evaluate the effect of the error of digitization using a navigated probe on the accuracy of the predicted position and orientation of an acetabular cup. An accuracy of approximately 0.3 mm resulted in a potential error of 1.5 mm in predicted cup position and 0.2 in inclination. It is possible to estimate expected accuracy through the use of statistical methods and, as a result, to optimise procedures and test methods using this knowledge.
引用
收藏
页码:S39 / S42
页数:4
相关论文
共 7 条
[1]  
*3M DEUTSCHL GMBH, 2005, FUNKT RETR
[2]  
Albertz J., 1989, PHOTOGRAMMETRIC GUID, p[60, 162]
[3]  
*AXIOS 3D SERV GMB, 2006, PROD INF
[4]  
LUHMANN T, 2006, CLOSE RANGE PHOTOGRA, P274
[5]  
LUTZ M, 2000, INTERAKTIVE MATH BEI
[6]  
*NO DIG INC, 2006, PROD INF COMP POL SY, P103
[7]  
*PLUS ORTH, 2005, PROD INF PIGALILEO C, P29